In vitro steady-flow analysis of systemic-to-pulmonary shunt haemodynamics.

G Pennati1, G B Fiore, F Migliavacca, K Laganà, R Fumero, G Dubini.   

Abstract

A modified Blalock-Taussig shunt is a connection created between the systemic and pulmonary arterial circulations to improve pulmonary perfusion in children with congenital heart diseases. Survival of these patients is critically dependent on blood flow distribution between the pulmonary and systemic circulations which in turn depends upon the flow resistance of the shunt. Previously, we investigated the pressure-flow relationship in rigid shunts with a computational approach. to estimate the pulmonary blood flow rate on the basis of the in vivo measured pressure drop. The present study aims at evaluating, in vitro how the anastomotic distensibility and restrictions due to suture presence affect the shunt pressure-flow relationship. Two actual Gore-Tex shunts (3 and 4 mm diameters) were sutured to compliant conduits by a surgeon and tested at different steady flow rates (0.25-11 min(-1)) and pulmonary pressures (3-34 mmHg). Corresponding computational models were also created to investigate the role of the anastomotic restrictions due to sutures. In vitro experiments showed that pulmonary artery pressure affects the pressure-flow relationship of the anastomoses. particularly at the distal site. However, this occurrence scarcely influences the total shunt pressure drop. Comparisons between in vitro and computational models without anastomotic restrictions show that the latter underestimates the in vitro pressure drops at any flow rate. The addition of the anastomotic restrictions (31 and 47% of the original area of 3 and 4 mm shunts, respectively) to the computational models reduces the gap, especially at high shunt flow rate and high pulmonary pressure.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11425077     DOI: 10.1016/s0021-9290(00)00167-6

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  5 in total

1.  Simulations reveal adverse hemodynamics in patients with multiple systemic to pulmonary shunts.

Authors:  Mahdi Esmaily-Moghadam; Bari Murtuza; Tain-Yen Hsia; Alison Marsden
Journal:  J Biomech Eng       Date:  2015-01-29       Impact factor: 2.097

2.  Hemodynamic analysis of sequential graft from right coronary system to left coronary system.

Authors:  Wenxin Wang; Boyan Mao; Haoran Wang; Xueying Geng; Xi Zhao; Huixia Zhang; Jinsheng Xie; Zhou Zhao; Bo Lian; Youjun Liu
Journal:  Biomed Eng Online       Date:  2016-12-28       Impact factor: 2.819

Review 3.  Computational fluid dynamics models and congenital heart diseases.

Authors:  Giancarlo Pennati; Chiara Corsini; Tain-Yen Hsia; Francesco Migliavacca
Journal:  Front Pediatr       Date:  2013-02-26       Impact factor: 3.418

Review 4.  Modeling single ventricle physiology: review of engineering tools to study first stage palliation of hypoplastic left heart syndrome.

Authors:  Giovanni Biglino; Alessandro Giardini; Tain-Yen Hsia; Richard Figliola; Andrew M Taylor; Silvia Schievano
Journal:  Front Pediatr       Date:  2013-10-30       Impact factor: 3.418

5.  3D Simulation Analysis of Central Shunt in Patient-Specific Hemodynamics: Effects of Varying Degree of Pulmonary Artery Stenosis and Shunt Diameters.

Authors:  Jiawei Liu; Haiyun Yuan; Neichuan Zhang; Xiangyu Chen; Chengbin Zhou; Meiping Huang; Qifei Jian; Jian Zhuang
Journal:  Comput Math Methods Med       Date:  2020-02-14       Impact factor: 2.238

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.