Use of an automated device for alternative site blood glucose monitoring.

S E Fineberg1, R M Bergenstal, R M Bernstein, L M Laffel, S L Schwartz.   

Abstract

OBJECTIVE: To evaluate the accuracy, comfort, and ease of use of a new automated device for blood glucose monitoring using the arm as an alternative sampling site. RESEARCH DESIGN AND METHODS: These studies use an automated hand-held device that applies a small vacuum, lances the skin, transfers blood onto an electrochemical test strip, and measures glucose. Patients who had type 1 or type 2 diabetes and had received no prior training using this device were recruited from five diabetes clinics. Testing was performed by the patients using this device and by trained healthcare professionals. Blood glucose was measured by 354 patients: from the arm using the device, from the finger using a laboratory reference instrument, and from the finger using the device via the secondary test port. Each patient completed a questionnaire rating the level of pain and ease of use of the device.
RESULTS: Blood glucose results in samples obtained from the arm with the automated device agreed well with finger-stick plasma glucose results using a reference instrument (regression slope 0.98, intercept 0.01 mmol/l [0.1 mg/dl], r = 0.96). Error grid analysis showed that 100% of the measurements fell within zones A and B. In the survey, 60% of the patients reported that arm testing with the automated device was "painless;" another 31% of the patients stated that it was "much less painful," and 6% of patients considered using the device "less painful" than finger-stick testing. In a survey containing 15 questions for rating the ease of use with a scale of 1 to 6, the overall mean rating was 5.5.
CONCLUSIONS: The automated device is easy to use and provides accurate glucose results; 97% of the patients found it less painful than finger-stick testing.

Entities:  

Mesh:

Year:  2001        PMID: 11423505     DOI: 10.2337/diacare.24.7.1217

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  5 in total

1.  Self-monitoring of blood glucose with finger tip versus alternative site sampling: effect on glycemic control in insulin-using patients with type 2 diabetes.

Authors:  Philip E Knapp; Kara M Showers; Jenna C Phipps; Jeanne L Speckman; Elliot Sternthal; Karen M Freund; Arlene S Ash; Caroline M Apovian
Journal:  Diabetes Technol Ther       Date:  2009-04       Impact factor: 6.118

Review 2.  New technologies and therapeutic approaches for the management of pediatric diabetes.

Authors:  M T Lawlor; L M Laffel
Journal:  Curr Diab Rep       Date:  2001-08       Impact factor: 4.810

3.  Forearm blood glucose testing in diabetes mellitus.

Authors:  S Greenhalgh; S Bradshaw; C M Hall; D A Price
Journal:  Arch Dis Child       Date:  2004-06       Impact factor: 3.791

4.  Impact of Squeezing Condition to Obtain Sufficient Blood Volume for Self-Monitoring of Blood Glucose Using an Automatic Puncturing and Sampling System.

Authors:  Mari Abe-Doi; Makoto Oe; Ryoko Murayama; Yasunobu Zushi; Hidenori Tanabe; Hiromi Sanada
Journal:  J Diabetes Sci Technol       Date:  2018-09

Review 5.  2D Materials in Development of Electrochemical Point-of-Care Cancer Screening Devices.

Authors:  Mohsen Mohammadniaei; Huynh Vu Nguyen; My Van Tieu; Min-Ho Lee
Journal:  Micromachines (Basel)       Date:  2019-09-30       Impact factor: 2.891

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.