Abstract
Kits for self-diagnosis of malaria by travelers in remote areas have been advocated. Antigen-capture test cards work reliably in the laboratory, but there is limited data regarding self-testing by ill patients. One hundred and fifty-three symptomatic patients presenting for a malaria test attempted self-diagnosis using these kits. Nine percent failed to perform a valid test. Against blood-film proven malaria, specificity was 97% (95% confidence intervals [CI] 93-99%) and sensitivity 95% (95% CI 74-99%). Although these results are encouraging, technical problems need to be addressed and test sensitivity proven by additional field-testing before this technology can be marketed to travelers.Entities:
Mesh:
Substances:
Year: 2000 PMID: 11421381
Source DB: PubMed Journal: Am J Trop Med Hyg ISSN: 0002-9637 Impact factor: 2.345