W U Laub1, A Bakai, F Nüsslin.
Abstract
The present study investigates the application of compensators for the intensity modulated irradiation of a thorax phantom. Measurements are compared with Monte Carlo and standard pencil beam algorithm dose calculations. Compensators were manufactured to produce the intensity profiles that were generated from the scientific version of the KonRad IMRT treatment-planning system for a given treatment plan. The comparison of dose distributions calculated with a pencil beam algorithm, with the Monte Carlo code EGS4 and with measurements is presented. By measurements in a water phantom it is demonstrated that the method used to manufacture the compensators reproduces the intensity profiles in a suitable manner. Monte Carlo simulations in a water phantom show that the accelerator head model used for simulations is sufficient. No significant overestimations of dose values inside the target volume by the pencil beam algorithm are found in the thorax phantom. An overestimation of dose values in lung by the pencil beam algorithm is also not found. Expected dose calculation errors of the pencil beam algorithm are suppressed, because the dose to the low density region lung is reduced by the use of a non-coplanar beam arrangement and by intensity modulation.Entities:
Mesh:
Year: 2001 PMID: 11419628 DOI: 10.1088/0031-9155/46/6/308