Intratester and intertester reproducibility of the lumbar motion monitor as a measure of range, velocity and acceleration of the thoracolumbar spine.

K P Gill1, M J Callaghan.   

Abstract

OBJECTIVE: The purpose of this study was to examine the intra- and intertester reproducibility of the Lumbar Motion Monitor (LMM) as a measure of thoracolumbar range of motion (RoM), velocity and acceleration.
DESIGN: The study was a clinical trial using asymptomatic subjects.
BACKGROUND: The LMM is a promising device from a clinical and research perspective, though the reproducibility of it's measurements has yet to be fully determined on subjects under free motion conditions.
METHODS: For intratester reproducibility, 15 subjects were required to move as far and as fast as they could in 8 s through flexion, side flexion, and rotation. They were tested on three occasions with 48 h separating tests. For intertester reproducibility, 10 subjects were tested by two examiners, and were required to move as far and as fast as they could in 8 s.
RESULTS: Intratester reproducibility coefficients ranged between 0.82 and 0.87 for RoM, 0.61 and 0.87 for velocity, and 0.46 and 0.72 for acceleration. Rotation had the poorest reproducibility in each instance. Intertester reproducibility ranged between 0.93 and 0.98.
CONCLUSIONS: The reproducibility of the LMM is suitably high for RoM and velocity for the device to be used for evaluation in a clinical and research setting.

Year:  1996        PMID: 11415654     DOI: 10.1016/0268-0033(96)00031-9

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

1.  Comparison of three validated systems to analyse spinal shape and motion.

Authors:  Bettina Dreischarf; Esther Koch; Marcel Dreischarf; Hendrik Schmidt; Matthias Pumberger; Luis Becker
Journal:  Sci Rep       Date:  2022-06-17       Impact factor: 4.996

2.  The effects of the Mulligan Sustained Natural Apophyseal Glide (SNAG) mobilisation in the lumbar flexion range of asymptomatic subjects as measured by the Zebris CMS20 3-D motion analysis system.

Authors:  Maria Moutzouri; Evdokia Billis; Nikolaos Strimpakos; Polixeni Kottika; Jacqueline A Oldham
Journal:  BMC Musculoskelet Disord       Date:  2008-10-01       Impact factor: 2.362

3.  Reliability and measurement error of frontal and horizontal 3D spinal motion parameters in 219 patients with chronic low back pain.

Authors:  Steen Harsted; Rune M Mieritz; Gert Bronfort; Jan Hartvigsen
Journal:  Chiropr Man Therap       Date:  2016-04-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.