Literature DB >> 11408127

Collamer intraocular lens: clinical results from the US FDA core study.

D C Brown1, S L Ziémba.   

Abstract

PURPOSE: To assess the safety and efficacy of the Staar Collamer intraocular lens (IOL) 1 year postoperatively.
SETTING: Fifteen private ophthalmology practices geographically distributed throughout the United States.
METHODS: Six hundred eighty-six cases with a mean age of 72.1 years were enrolled in a 2-phase U.S. Food and Drug Administration (FDA) clinical study of the Collamer IOL implanted during phacoemulsification for cataract removal. The 12 month follow-up comprised data from 502 cases.
RESULTS: Twelve months postoperatively, 96.2% of all cases and 95.6% without preexisting pathology achieved 20/40 or better best corrected visual acuity. Both results were better than FDA Grid values. No persistent sight-threatening complications, IOL dislocations, or IOL removals were reported.
CONCLUSIONS: Results show that the Collamer IOL is safe and effective for use in small incision cataract surgery.

Entities:  

Mesh:

Year:  2001        PMID: 11408127     DOI: 10.1016/s0886-3350(01)00785-4

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  A critical evaluation of longitudinal changes of astigmatism following implantation of toric implantable collamer lens (TICL): a comparison between treated and untreated cases over 4 years.

Authors:  Maja Bohac; Alma Biscevic; Ivan Gabric; Kresimir Gabric; Violeta Shijakova; Sudi Patel
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-10-16       Impact factor: 3.117

2.  Traumatic dislocation of implanted collamer phakic lens: a case report and review of the literature.

Authors:  Majid Moshirfar; Brian C Stagg; Valliammai Muthappan; Shail A Vasavada
Journal:  Open Ophthalmol J       Date:  2014-05-30
  2 in total