Evaluation of i-STAT creatinine assay.

M H Gault1, M E Seymour, W E Howell.   

Abstract

BACKGROUND: We evaluated a new, 2-min blood creatinine method using the hand-held i-STAT analyzer. Good results have already been reported using this analyzer for 10 methods including electrolytes, TCO2, pH, PCO2, bicarbonate, glucose, hemoglobin and urea for uremic blood, hemodialysate and peritoneal effluent.
METHODS: Evaluation included study of imprecision and accuracy.
RESULTS: Imprecision studies gave excellent results, including those for reproducibility of 6 solutions with a mean of 10 repeats and coefficients of variation (CVs) of 0.4-3.4%, and also the mean of the differences between 33 duplicate blood specimens which was 2.2% of the specimen mean. To assess accuracy, we compared results of 149 tests by i-STAT and Beckman Synchron CX7 methods. The difference between the two means was 2.6% and the mean of all differences was 10.9% with i-STAT results higher, especially when blood creatinine values were < 100 micromol/l (1.1 mg/dl) indicating the need for a slightly higher upper limit of the normal range. The correlation coefficient between the two methods was 0.99, the slope 1.0 and the intercept -5.0 micromol/l (-0.06 mg/dl). We assessed the recommended creatinine correction for variation in PCO2 above and below 40 mm Hg, but our results did not suggest the need for such a correction in our range of 27-64 mm Hg; omission would remove a major method disadvantage. Assays of hemodialysate and peritoneal effluent were also satisfactory.
CONCLUSIONS: The i-STAT creatinine method is simple and rapid and our evaluation showed satisfactory accuracy and precision. However, results were on average slightly higher than for the Beckman Synchron CX7 method. Copyright 2001 S. Karger AG, Basel

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11399924     DOI: 10.1159/000045982

Source DB:  PubMed          Journal:  Nephron        ISSN: 1660-8151            Impact factor:   2.847


  3 in total

1.  Serum creatinine detection by a conducting-polymer-based electrochemical sensor to identify allograft dysfunction.

Authors:  Fang Wei; Scott Cheng; Yael Korin; Elaine F Reed; David Gjertson; Chih-ming Ho; H Albin Gritsch; Jeffrey Veale
Journal:  Anal Chem       Date:  2012-08-24       Impact factor: 6.986

2.  Point-of-care creatinine tests to assess kidney function for outpatients requiring contrast-enhanced CT imaging: systematic reviews and economic evaluation.

Authors:  Mark Corbett; Ana Duarte; Alexis Llewellyn; James Altunkaya; Melissa Harden; Martine Harris; Simon Walker; Stephen Palmer; Sofia Dias; Marta Soares
Journal:  Health Technol Assess       Date:  2020-08       Impact factor: 4.014

3.  Performance evaluation of all analytes on the epoc® Blood Analysis System for use in hospital surgical and intensive care units.

Authors:  Zahraa Mohammed-Ali; Sousan Bagherpoor; Pauline Diker; Thuy Hoang; Ivana Vidovic; Christine Cursio; Felix Leung; Davor Brinc
Journal:  Pract Lab Med       Date:  2020-11-25
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.