Interlaboratory reproducibility of standard accelerated aging methods for oxidation of UHMWPE.

S M Kurtz1, O K Muratoglu, F Buchanan, B Currier, R Gsell, K Greer, G Gualtieri, R Johnson, S Schaffner, K Sevo, S Spiegelberg, F W Shen, S S Yau.   

Abstract

During accelerating aging, experimental uncertainty may arise due to variability in the oxidation process, or due to limitations in the technique that is ultimately used to measure oxidation. The purpose of the present interlaboratory study was to quantify the repeatability and reproducibility of standard accelerated aging methods for ultra-high molecular weight polyethylene (UHMWPE). Sections (200 microm thick) were microtomed from the center of an extruded rod of GUR 4150 HP, gamma irradiated in air or nitrogen, and circulated to 12 institutions in the United States and Europe for characterization of oxidation before and after accelerated aging. Specimens were aged for 3 weeks at 80 degrees C in an air circulating oven or for 2 weeks at 70 degrees C in an oxygen bomb (maintained at 503 kPa (5 atm.) of O2) in accordance with the two standard protocols described in ASTM F 2003-00. FTIR spectra were collected from each specimen within 24 h of the start and finish of accelerated aging, and oxidation indices were calculated by normalizing the peak area of the carbonyl region by the reference peak areas at 1370 or 2022 cm(-1). The mean relative interlaboratory uncertainty of the oxidation data was 78.5% after oven aging and 129.1% after bomb aging. The oxidation index measurement technique was not found to be a significant factor in the reproducibility. Comparable relative intrainstitutional uncertainty was observed after oven aging and bomb aging. For both aging methods, institutions successfully discriminated between air-irradiated and control specimens. However, the large interinstitutional variation suggests that absolute performance standards for the oxidation index of UHMWPE after accelerated aging may not be practical at the present time.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11396876     DOI: 10.1016/s0142-9612(00)00333-1

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  3 in total

1.  On the assessment of oxidative and microstructural changes after in vivo degradation of historical UHMWPE knee components by means of vibrational spectroscopies and nanoindentation.

Authors:  Francisco J Medel; Clare M Rimnac; Steven M Kurtz
Journal:  J Biomed Mater Res A       Date:  2009-05       Impact factor: 4.396

2.  Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions.

Authors:  Carmen Zietz; Joern Reinders; Jens Schwiesau; Alexander Paulus; Jan Philippe Kretzer; Thomas Grupp; Sandra Utzschneider; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2015-02-26       Impact factor: 3.896

3.  A comparison of the efficacy of various antioxidants on the oxidative stability of irradiated polyethylene.

Authors:  Natalie Hope; Anuj Bellare
Journal:  Clin Orthop Relat Res       Date:  2015-03       Impact factor: 4.176

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.