Skin topography measurement by interference fringe projection: a technical validation.

J M Lagarde1, C Rouvrais, D Black, S Diridollou, Y Gall.   

Abstract

BACKGROUND/AIMS: The quantitative analysis of skin topography is frequently used in cosmetology to evaluate the efficacy of hydrating or anti-wrinkle creams (micro-topography studies) or creams for slimming or to alleviate stretch marks (macro-topography studies). Numerous methods involving the three-dimensional (3-D) reconstruction of the topography from silicone replicas have been developed. Some of the main techniques applied include optical profilometry based on the reflection of oblique lighting or on transparency, and profilometry by laser focusing or triangulation.
METHODS: We chose to test the Dermatop system developed by Eotech Co., which is based on interference fringe projection profilometry associated with the Toposurf surface processing software. Owing to the system's principle of operation, we first ran three experiments to evaluate the influence of ambient lighting on the 3-D reconstruction and the surface roughness parameters, the influence of the calibration device built into the acquisition system and the influence of calibration in terms of ambient lighting correction. We then evaluated the accuracy, the repeatability and the reproducibility of the measurements provided by the system from known metal topographic standards and skin replicas.
RESULTS: It was seen that external lighting can cause variations in the calculation of parameters SPa, SPt and SPtm which can reach error levels of about 5% if the operator does not almost systematically calibrate the apparatus between each measurement. These three experiments enabled us to standardise the protocol for the acquisition of 3-D information, in order to minimise the lighting problems and yet respect the requirements for routine studies. The correlation coefficient between the theoretical values and the mean of experimental values was 0.9955 for parameter SPa and 0.9983 for parameter SPtm. Repeatability presented variations of under 4%, irrespective of the parameters, and reproducibility of under 6%.
CONCLUSIONS: The technique provides very satisfactory results from the point of view of accuracy, repeatability and reproducibility. It is an excellent compromise between cost, accuracy and the time required.

Entities:  

Mesh:

Year:  2001        PMID: 11393203     DOI: 10.1034/j.1600-0846.2001.70210.x

Source DB:  PubMed          Journal:  Skin Res Technol        ISSN: 0909-752X            Impact factor:   2.365


  7 in total

1.  Prediction of Android and Gynoid Body Adiposity via a Three-dimensional Stereovision Body Imaging System and Dual-Energy X-ray Absorptiometry.

Authors:  Jane J Lee; Jeanne H Freeland-Graves; M Reese Pepper; Philip R Stanforth; Bugao Xu
Journal:  J Am Coll Nutr       Date:  2015-04-27       Impact factor: 3.169

2.  Ultrasound and 3D Skin Imaging: Methods to Evaluate Efficacy of Striae Distensae Treatment.

Authors:  Mariella Bleve; Priscilla Capra; Franca Pavanetto; Paola Perugini
Journal:  Dermatol Res Pract       Date:  2011-11-22

3.  Design, characterization, and clinical evaluation of argan oil nanostructured lipid carriers to improve skin hydration.

Authors:  Deise Michele Tichota; Ana Catarina Silva; José Manuel Sousa Lobo; Maria Helena Amaral
Journal:  Int J Nanomedicine       Date:  2014-08-11

4.  Redefining face contour with a novel anti-aging cosmetic product: an open-label, prospective clinical study.

Authors:  Aurora Garre; Gemma Martinez-Masana; Jaime Piquero-Casals; Corinne Granger
Journal:  Clin Cosmet Investig Dermatol       Date:  2017-11-13

Review 5.  Optical Technologies for the Improvement of Skin Cancer Diagnosis: A Review.

Authors:  Laura Rey-Barroso; Sara Peña-Gutiérrez; Carlos Yáñez; Francisco J Burgos-Fernández; Meritxell Vilaseca; Santiago Royo
Journal:  Sensors (Basel)       Date:  2021-01-02       Impact factor: 3.576

Review 6.  A systematic review of objective burn scar measurements.

Authors:  Kwang Chear Lee; Janine Dretzke; Liam Grover; Ann Logan; Naiem Moiemen
Journal:  Burns Trauma       Date:  2016-04-27

7.  A serum containing vitamins C & E and a matrix-repair tripeptide reduces facial signs of aging as evidenced by Primos® analysis and frequently repeated auto-perception.

Authors:  Karl Lintner; Francine Gerstein; Nowell Solish
Journal:  J Cosmet Dermatol       Date:  2020-10-26       Impact factor: 2.696

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.