Reduction of oscillatory pressure along the endotracheal tube is indicative for maximal respiratory compliance during high-frequency oscillatory ventilation: a mathematical model study.

H R van Genderingen1, A Versprille, T Leenhoven, D G Markhorst, A J van Vught, R M Heethaar.   

Abstract

We hypothesized that during high-frequency oscillatory ventilation (HFOV), a reduction of peak-to-peak oscillatory pressure along the endotracheal tube is maximal when respiratory system compliance is maximal. We made a mathematical model of the endotracheal tube and the respiratory system of a neonate suffering from idiopathic respiratory distress syndrome (IRDS). The model consisted of linear viscous and inertive elements, a non-linear endotracheal tube resistance, and a non-linear compliance allowing for alveolar recruitment and overdistention. Respiratory compliance was maximal at the transition between maximal recruitment and minimal overdistention. A new variable, the oscillatory pressure ratio (OPR), was defined as the ratio between peak-to-peak oscillatory pressures at the distal end and the proximal opening of the endotracheal tube, respectively. The respiratory variables of four patients were fed into the model, and the relationship between respiratory system compliance and OPR was determined. OPR decreased as compliance increased, except for very low compliances below where 0.08 mL. cm H2O(-1), and OPR increased with increasing compliance. The relationship between mean airway pressure P(aw) and OPR revealed that the minimal OPR (range, 0.37-0.78) and maximal respiratory compliance coincided at the same P(aw). However, the relationship did depend on oscillation frequency, applied oscillatory pressure, and endotracheal tube resistance, parameters that may change during clinical application of HFOV. When 81 permutations of nominal and extreme respiratory variables were used in the model, the minimum OPR (0.60 +/- 0.23) and maximum compliance coincided in all cases. These model experiments support our hypothesis. The results indicate that the OPR may be a useful index to optimize lung expansion, where lung recruitment is maximal and overdistention minimal. In vivo tests will be needed to reveal the feasibility and reliability of such an index for biomedical and clinical application.

Entities:  

Mesh:

Year:  2001        PMID: 11389579     DOI: 10.1002/ppul.1075

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  4 in total

1.  Breath-to-breath analysis of abdominal and rib cage motion in surfactant-depleted piglets during high-frequency oscillatory ventilation.

Authors:  Dick G Markhorst; Jos R C Jansen; Adrianus J van Vught; Huibert R van Genderingen
Journal:  Intensive Care Med       Date:  2005-01-20       Impact factor: 17.440

Review 2.  [High-frequency oscillatory ventilation. Ventilation procedure for adults with acute lung failure].

Authors:  M David; C Werner
Journal:  Anaesthesist       Date:  2007-05       Impact factor: 1.041

3.  Effect of frequency on pressure cost of ventilation and gas exchange in newborns receiving high-frequency oscillatory ventilation.

Authors:  Emanuela Zannin; Raffaele L Dellaca'; Giulia Dognini; Lara Marconi; Martina Perego; Jane J Pillow; Paolo E Tagliabue; Maria Luisa Ventura
Journal:  Pediatr Res       Date:  2017-07-26       Impact factor: 3.756

Review 4.  The Physiological Basis of High-Frequency Oscillatory Ventilation and Current Evidence in Adults and Children: A Narrative Review.

Authors:  Andrew G Miller; Herng Lee Tan; Brian J Smith; Alexandre T Rotta; Jan Hau Lee
Journal:  Front Physiol       Date:  2022-04-26       Impact factor: 4.755

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.