Literature DB >> 11388569

Adsorption of radiopharmaceuticals to syringes leads to lower administered activity than intended.

R D Gunasekera1, A Notghi, A B Mostafa, L K Harding.   

Abstract

In this study we examined the delivered activity of radiopharmaceuticals injected into patients and the various factors which lead to incomplete delivery. Four commonly used 99Tcm radiopharmaceuticals were included, namely pertechnetate, methylene diphosphonate (MDP), macro-aggregated albumin (MAA) and tetrofosmin. There was no appreciable residual activity after injection of pertechnetate and MDP other than that due to the dead space in syringes (0.06 ml, 4% of measured activity). However, there was considerable residual activity following an MAA injection. This was particularly dependent on the radiopharmaceutical residence time in the syringe prior to injection. If MAA was pre-dispensed there was up to 80% of activity remaining in the syringe following injection. We have shown that this activity is not bound and can be decreased by agitating the syringe prior to the injection. For tetrofosmin, the activity remaining in the syringe was shown to be mostly due to adsorption of the tetrofosmin onto the plastic syringe (up to 30% of the activity); in particular, onto the rubber plunger. The amount of binding probably depends on the formulation of the rubber plunger and the plastic barrel, which varies between manufacturers. Of the three commonly used syringe brands (Becton-Dickinson, Sherwood, and Braun syringes) the Braun syringes were shown to have the least binding (6%) and would be most suitable for tetrofosmin injection. We recommend the use of Braun syringes for tetrofosmin. We also recommend that pre-dispensed MAA injections should be gently agitated before injection into the patient. No special precautions are necessary for pertechnetate or MDP injections.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11388569     DOI: 10.1097/00006231-200105000-00005

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  6 in total

1.  The clinical importance of accurate measurement of injected doses for radionuclide myocardial perfusion imaging.

Authors:  Raymond Taillefer
Journal:  J Nucl Cardiol       Date:  2015-08-26       Impact factor: 5.952

2.  Adherence of Tc-99m agents to plastic syringes.

Authors:  Trifon Spyridonidis; Theodoros Skouras; Dimitris J Apostolopoulos
Journal:  J Nucl Cardiol       Date:  2016-09-07       Impact factor: 5.952

3.  99mTc-sestamibi using a direct conversion molecular breast imaging system to assess tumor response to neoadjuvant chemotherapy in women with locally advanced breast cancer.

Authors:  David Mitchell; Carrie B Hruska; Judy C Boughey; Dietlind L Wahner-Roedler; Katie N Jones; Cindy Tortorelli; Amy Lynn Conners; Michael K O'Connor
Journal:  Clin Nucl Med       Date:  2013-12       Impact factor: 7.794

4.  Adherence of Tc-99 Sestamibi to Plastic Syringes Could Complicate Efforts in Dose Reduction in MPI SPECT.

Authors:  Sean N Reynolds; Janusz Kikut
Journal:  J Nucl Cardiol       Date:  2015-08-26       Impact factor: 5.952

Review 5.  Breast-Specific Gamma Imaging: An Added Value in the Diagnosis of Breast Cancer, a Systematic Review.

Authors:  Maria Silvia De Feo; Marko Magdi Abdou Sidrak; Miriam Conte; Viviana Frantellizzi; Andrea Marongiu; Flaminia De Cristofaro; Susanna Nuvoli; Angela Spanu; Giuseppe De Vincentis
Journal:  Cancers (Basel)       Date:  2022-09-23       Impact factor: 6.575

6.  Unpredictable Nanoparticle Retention in Commonly Used Plastic Syringes Introduces Dosage Uncertainties That May Compromise the Accuracy of Nanomedicine and Nanotoxicology Studies.

Authors:  Uwe Holzwarth; Unai Cossío; Jordi Llop; Wolfgang G Kreyling
Journal:  Front Pharmacol       Date:  2019-11-06       Impact factor: 5.810

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.