P Ritz1. 1. Centre Hospitalier Universitaire, Angers, France. patrick.ritz@wanadoo.fr
Abstract
BACKGROUND: This study validates, in geriatric patients, bioelectrical impedance analysis (BIA) equations that had been derived to estimate total body water (TBW) and extracellular water (ECW) in healthy elderly subjects. METHODS: We performed a multicentric trial in six geriatric wards. We studied 169 patients with varying degrees of hydration: dehydrated, euvolemic, and overhydrated. BIA estimates of TBW and of ECW were compared with the measurement of TBW with (18)O dilution and of ECW with bromide (Br) dilution. RESULTS: BIA estimated TBW with a difference of 0.48 +/- 2.3 l (mean +/- SD) (50 kHz; p = .01) and 0.69 +/- 2.2 l (100 kHz; p < 0.001) compared with (18)O dilution. The difference was not affected by the hydration status. Estimates of ECW with BIA were systematically biased compared with Br dilution: 4.6 +/- 3.1 l (equation from Segal and colleagues; p < .001) and 3.4 +/- 2.9 l (equation from Visser and colleagues; p < .001). We propose a new, cross-validated equation. Conclusions. Body water spaces can be estimated accurately in geriatric patients with BIA.
BACKGROUND: This study validates, in geriatric patients, bioelectrical impedance analysis (BIA) equations that had been derived to estimate total body water (TBW) and extracellular water (ECW) in healthy elderly subjects. METHODS: We performed a multicentric trial in six geriatric wards. We studied 169 patients with varying degrees of hydration: dehydrated, euvolemic, and overhydrated. BIA estimates of TBW and of ECW were compared with the measurement of TBW with (18)O dilution and of ECW with bromide (Br) dilution. RESULTS: BIA estimated TBW with a difference of 0.48 +/- 2.3 l (mean +/- SD) (50 kHz; p = .01) and 0.69 +/- 2.2 l (100 kHz; p < 0.001) compared with (18)O dilution. The difference was not affected by the hydration status. Estimates of ECW with BIA were systematically biased compared with Br dilution: 4.6 +/- 3.1 l (equation from Segal and colleagues; p < .001) and 3.4 +/- 2.9 l (equation from Visser and colleagues; p < .001). We propose a new, cross-validated equation. Conclusions. Body water spaces can be estimated accurately in geriatric patients with BIA.
Authors: James S Powers; Leena Choi; Rhonda Bitting; Nitin Gupta; Maciej Buchowski Journal: J Gerontol A Biol Sci Med Sci Date: 2009-02-19 Impact factor: 6.053
Authors: Lee Hooper; Asmaa Abdelhamid; Natalie J Attreed; Wayne W Campbell; Adam M Channell; Philippe Chassagne; Kennith R Culp; Stephen J Fletcher; Matthew B Fortes; Nigel Fuller; Phyllis M Gaspar; Daniel J Gilbert; Adam C Heathcote; Mohannad W Kafri; Fumiko Kajii; Gregor Lindner; Gary W Mack; Janet C Mentes; Paolo Merlani; Rowan A Needham; Marcel G M Olde Rikkert; Andreas Perren; James Powers; Sheila C Ranson; Patrick Ritz; Anne M Rowat; Fredrik Sjöstrand; Alexandra C Smith; Jodi J D Stookey; Nancy A Stotts; David R Thomas; Angela Vivanti; Bonnie J Wakefield; Nana Waldréus; Neil P Walsh; Sean Ward; John F Potter; Paul Hunter Journal: Cochrane Database Syst Rev Date: 2015-04-30