| Literature DB >> 11382427 |
P V Giannoudis1, S J Matthews, R M Smith.
Abstract
The development of solid nails has not eliminated the risk of occasional nail failure. It has been suggested that subsequent intra-medullary removal of solid nail fragment may be virtually impossible and thus major surgery is required for their revision. We report two cases of hardware failure of a solid intra-medullary nail (Synthes solid femoral and tibial nail) and describe their successful intra-medullary removal using the Synthes extraction kit.Mesh:
Year: 2001 PMID: 11382427 DOI: 10.1016/s0020-1383(00)00236-9