[Comparative biomechanical compression trials with a new vertebral prosthetic implant].

C Knop1, U Lange, L Bastian, M Oeser, M Blauth.   

Abstract

UNLABELLED: The authors present a new titanium implant for replacement of the vertebral body (Synex). Possible indications would be fractures or dislocations with destruction of the anterior column, posttraumatic kyphosis as well as tumors in the throracolumbar spine. The construction has to be completed by a stabilizing implant. For best fit and contact to adjacent end-plates Synex is distractable in situ. The possibility of secondary dislocation or loss of correction should thereby be minimised.
OBJECTIVES: We performed comparative compression tests with Synex and MOSS ("Harms mesh cage") on human cadaveric specimens of intact vertebrae (L1). The aim of the study was to measure the compressive strength of the vertebral body end-plate in uniaxial loading via both implants to exclude a caving of Synex in vivo.
METHODS: 12 human cadaveric specimens of intact vertebrae (L1) were divided in 2 similar groups (matched pairs) according to bone mineral density (BMD), determined using DE-QCT. The specimens were loaded with axial compression force at a constant speed of 5 mm/min to failure and the displacement was recorded with a continuous load-displacement curve.
RESULTS: The mean ultimate compression force (Fmax) showed a tendency towards a higher result testing Synex with 3396 N versus 2719 N (non significant). The displacement until Fmax was 2.9 mm in group S (Synex), which was half as long as in group M (5.8 mm). The difference was significant (p < 0.001). The compression force was twice as high and significantly (p < 0.05) higher with Synex at a displacement of 1 mm, 1.5 mm and 2 mm. A significant (p < 0.001) correlation (R = 0.89) between Fmax and BMD was found.
CONCLUSIONS: Synex was found to be at least comparable to MOSS for suspensory replacement of the vertebral body at the thoracolumbar spine. A possible consequence of the significantly higher mean compression forces between 1 and 2 mm displacement might be a decreased segmental deformation or loss of correction.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11381759     DOI: 10.1007/s001130050684

Source DB:  PubMed          Journal:  Unfallchirurg        ISSN: 0177-5537            Impact factor:   1.000


  2 in total

1.  Primary stability of anterior lumbar stabilization: interdependence of implant type and endplate retention or removal.

Authors:  Christian H Flamme; Nadine von der Heide; Caroline Heymann; Christof Hurschler
Journal:  Eur Spine J       Date:  2005-08-10       Impact factor: 3.134

Review 2.  Endoscopic surgery on the thoracolumbar junction of the spine.

Authors:  Rudolf Beisse
Journal:  Eur Spine J       Date:  2006-02-11       Impact factor: 3.134

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.