Effectiveness of couch height-based patient set-up and an off-line correction protocol in prostate cancer radiotherapy.

E N van Lin1, E Nijenhuis, H Huizenga, L van der Vight, A Visser.   

Abstract

PURPOSE: To investigate set-up improvement caused by applying a couch height-based patient set-up method in combination with a technologist-driven off-line correction protocol in nonimmobilized radiotherapy of prostate patients. METHODS AND MATERIALS: A three-dimensional shrinking action level correction protocol is applied in two consecutive patient cohorts with different set-up methods: the traditional "laser set-up" group (n = 43) and the "couch height set-up" group (n = 112). For all directions, left-right, ventro-dorsal, and cranio-caudal, random and systematic set-up deviations were measured.
RESULTS: The couch height set-up method improves the patient positioning compared to the laser set-up method. Without application of the correction protocol, both systematic and random errors reduced to 2.2-2.4 mm (1 SD) and 1.7-2.2 mm (1 SD), respectively. By using the correction protocol, systematic errors reduced further to 1.3-1.6 mm (1 SD). One-dimensional deviations were within 5 mm for >90% of the measured fractions. The required number of corrections per patient in the off-line correction protocol was reduced significantly during the course of treatment from 1.1 to 0.6 by the couch height set-up method. The treatment time was not prolonged by application of the correction protocol.
CONCLUSIONS: The couch height set-up method improves the set-up significantly, especially in the ventro-dorsal direction. Combination of this set-up method with an off-line correction strategy, executed by technologists, reduces the number of set-up corrections required.

Entities:  

Mesh:

Year:  2001        PMID: 11380247     DOI: 10.1016/s0360-3016(01)01520-6

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  Verification of the dose attenuation of a newly developed vacuum cushion for intensity-modulated radiation therapy of prostate cancer.

Authors:  Toru Takakura; Yoshiyuki Ito; Akinori Higashikawa; Tomohiro Nishiyama; Takashi Sakamoto
Journal:  Radiol Phys Technol       Date:  2016-06-03

2.  Technical note: improved positioning protocol for patient setup accuracy in conventional radiotherapy for lung cancer.

Authors:  Hongbo Chai; Yuichiro Narita; Masafumi Takagi; Mikiko Kudo; Tomomi Kimura; Keiichi Kattou
Journal:  Radiol Phys Technol       Date:  2019-09-23

3.  Quality improvement process to assess tattoo alignment, set-up accuracy and isocentre reproducibility in pelvic radiotherapy patients.

Authors:  Kelly Elsner; Kate Francis; George Hruby; Stephanie Roderick
Journal:  J Med Radiat Sci       Date:  2014-11-21

4.  Camera-based independent couch height verification in radiation oncology.

Authors:  Martijn Kusters; Rob Louwe; Liesbeth Biemans-van Kastel; Henk Nieuwenkamp; Rien Zahradnik; Roy Claessen; Ronald van Seters; Henk Huizenga
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.