Literature DB >> 11379677

Automated responsiveness test (ART) predicts loss of consciousness and adverse physiologic responses during propofol conscious sedation.

A G Doufas1, M Bakhshandeh, A R Bjorksten, R Greif, D I Sessler.   

Abstract

BACKGROUND: The authors evaluated a device designed to provide conscious sedation with propofol (propofol-air), or propofol combined with 50% nitrous oxide (N2O; propofol-N2O). An element of this device is the automated responsiveness test (ART), a method for confirming that patients remain conscious. The authors tested the hypotheses that the ART predicts loss of consciousness and that failure to respond to the ART precedes sedation-induced respiratory or hemodynamic toxicity.
METHODS: The protocol consisted of sequential 15-min cycles in 20 volunteers. After a 15-min control period, propofol was infused to an initial target effect-site concentration of 0.0 microg/ml with N2O or 1.5 microg/ml with air. Subsequently, the propofol target effect-site concentration was increased by a designated increment (0.25 and 0.5 microg/ml) and the process repeated. This sequence was continued until loss of consciousness, as defined by an Observer's Assessment of Alertness/Sedation (OAA/S) score of 10/20 or less, or until an adverse physiologic event was detected.
RESULTS: The OAA/S score at which only 50% of the volunteers were able to respond to the ART (P50) during propofol-N2O was 11.1 of 20 (95% confidence interval [CI]: 10.6-11.8); the analogous P50 was 11.8 of 20 (95% CI: 11.4-12.3) with propofol-air. Failure to respond to the ART occurred at a plasma propofol concentration of 0.7 +/- 0.6 microg/ml with propofol-N2O and 1.6 +/- 0.6 microg/ml with propofol-air, whereas loss of consciousness occurred at 1.2 +/- 0.8 microg/ml and 1.9 +/- 0.7 microg/ml, respectively. There were no false-normal ART responses.
CONCLUSION: The ART can guide individual titration of propofol because failure to respond to responsiveness testing precedes loss of consciousness and is not susceptible to false-normal responses. The use of N2O with propofol for conscious sedation decreases the predictive accuracy of the ART.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11379677     DOI: 10.1097/00000542-200104000-00010

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  5 in total

1.  Induction speed is not a determinant of propofol pharmacodynamics.

Authors:  Anthony G Doufas; Maryam Bakhshandeh; Andrew R Bjorksten; Steven L Shafer; Daniel I Sessler
Journal:  Anesthesiology       Date:  2004-11       Impact factor: 7.892

2.  Propofol infusion for sedation during spinal anesthesia.

Authors:  Tomoki Nishiyama
Journal:  J Anesth       Date:  2007-05-30       Impact factor: 2.078

3.  Cortical functional connectivity indexes arousal state during sleep and anesthesia.

Authors:  Matthew I Banks; Bryan M Krause; Christopher M Endemann; Declan I Campbell; Christopher K Kovach; Mark Eric Dyken; Hiroto Kawasaki; Kirill V Nourski
Journal:  Neuroimage       Date:  2020-02-08       Impact factor: 6.556

4.  Individual effect-site concentrations of propofol are similar at loss of consciousness and at awakening.

Authors:  Hiroko Iwakiri; Noboru Nishihara; Osamu Nagata; Takashi Matsukawa; Makoto Ozaki; Daniel I Sessler
Journal:  Anesth Analg       Date:  2005-01       Impact factor: 6.627

5.  Relationship between UGT1A9 gene polymorphisms, efficacy, and safety of propofol in induced abortions amongst Chinese population: a population-based study.

Authors:  Ying-Bin Wang; Rong-Zhi Zhang; Sheng-Hui Huang; Shu-Bao Wang; Jian-Qin Xie
Journal:  Biosci Rep       Date:  2017-10-24       Impact factor: 3.840

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.