Mathematical coupling does not explain the relationship between right ventricular end-diastolic volume and cardiac output.

L D Nelson1, K Safcsak, M L Cheatham, E F Block.   

Abstract

OBJECTIVE: To evaluate the clinical significance of mathematical coupling on the correlation between cardiac output and right ventricular end-diastolic volume (RVEDV) through measurement of cardiac output by two independent techniques.
DESIGN: Prospective, observational study.
SETTING: Surgical intensive care unit in a level 1 trauma center. PATIENTS: Twenty-eight critically ill surgical patients who received mechanical ventilation and hemodynamic monitoring with a pulmonary artery catheter.
INTERVENTIONS: A pulmonary artery catheter designed to measure right ventricular ejection fraction (RVEF) and cardiac output by the intermittent bolus thermodilution (TDCO) method and continuous cardiac output by the pulsed thermal energy technique was placed. A computerized data logger was used to collect data simultaneously from the RVEF/TDCO system and the continuous cardiac output system.
MEASUREMENTS AND MAIN RESULTS: Two hundred forty-nine data sets from 28 patients were compared. There is statistical correlation between TDCO and continuous cardiac output measurements (r = 0.95, p < 0.0001) with an acceptable bias (-0.11 L/min) and precision (+/-0.74 L/min). The correlation was maintained over a wide range of cardiac outputs (2.3-17.8 L/min). There is a high degree of correlation between RVEDV and both TDCO (r = 0.72, p < 0.0001) and independently measured continuous cardiac output (r = 0.68, p < 0.0001). These correlation coefficients are not statistically different (p = 0.15).
CONCLUSIONS: The continuous cardiac output technique accurately approximates cardiac output measured by the TDCO method. RVEDV calculated from TDCO correlates well with both TDCO and independently measured continuous cardiac output. Because random measurement errors of the two techniques differ, mathematical coupling alone does not explain the correlation between RVEDV estimates of preload and cardiac output.

Entities:  

Mesh:

Year:  2001        PMID: 11378601     DOI: 10.1097/00003246-200105000-00006

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  3 in total

1.  Agreement between continuous and intermittent pulmonary artery thermodilution for cardiac output measurement in perioperative and intensive care medicine: a systematic review and meta-analysis.

Authors:  Karim Kouz; Frederic Michard; Alina Bergholz; Christina Vokuhl; Luisa Briesenick; Phillip Hoppe; Moritz Flick; Gerhard Schön; Bernd Saugel
Journal:  Crit Care       Date:  2021-03-29       Impact factor: 9.097

2.  Continuously assessed right ventricular end-diastolic volume as a marker of cardiac preload and fluid responsiveness in mechanically ventilated cardiac surgical patients.

Authors:  Christoph Wiesenack; Christoph Fiegl; Andreas Keyser; Sven Laule; Christopher Prasser; Cornelius Keyl
Journal:  Crit Care       Date:  2005-04-01       Impact factor: 9.097

Review 3.  The contemporary pulmonary artery catheter. Part 2: measurements, limitations, and clinical applications.

Authors:  I T Bootsma; E C Boerma; T W L Scheeren; F de Lange
Journal:  J Clin Monit Comput       Date:  2021-03-01       Impact factor: 2.502

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.