A new high intensity focused ultrasound applicator for surgical applications.

M D Brentnall1, R W Martin, S Vaezy, P Kaczkowski, F Forster, L Crum.   

Abstract

Improved high-intensity focused ultrasound (HIFU) surgical applicators are required for use in a surgical environment. We report on the performance and characteristics of a new solid-cone HIFU applicator. Previous HIFU devices used a water-filled stand-off to couple the ultrasonic energy from the transducer to the treatment area. The new applicator uses a spherically-focused element and a solid aluminum cone to guide and couple the ultrasound to the tissue. Compared with the water-filled applicators, this new applicator is simpler to set up and manipulate, cannot leak, prevents the possibility of cavitation within the coupling device, and is much easier to sterilize and maintain during surgery. The design minimizes losses caused by shear wave conversion found in tapered solid acoustic velocity transformers operated at high frequencies. Computer simulations predicted good transfer of longitudinal waves. Impedance measurements, beam plots, Schlieren images, and force balance measurements verified strong focusing and suitable transfer of acoustic energy into water. At the focus, the -3 dB beam dimensions are 1.2 mm (axial) x 0.3 mm (transverse). Radiation force balance measurements indicate a power transfer efficiency of 40%. In vitro and in vivo tissue experiments confirmed the applicator's ability to produce hemostasis.

Entities:  

Mesh:

Year:  2001        PMID: 11367806     DOI: 10.1109/58.895907

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  5 in total

1.  Experimental evaluation of lesion prediction modelling in the presence of cavitation bubbles: intended for high-intensity focused ultrasound prostate treatment.

Authors:  L Curiel; F Chavrier; B Gignoux; S Pichardo; S Chesnais; J Y Chapelon
Journal:  Med Biol Eng Comput       Date:  2004-01       Impact factor: 2.602

2.  Feasibility of MR-temperature mapping of ultrasonic heating from a CMUT.

Authors:  Serena H Wong; Ronald D Watkins; Mario Kupnik; Kim Butts Pauly; Butrus T Khuri-Yakub
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2008-04       Impact factor: 2.725

3.  High-intensity focused ultrasound ablation: an effective and safe treatment for secondary hypersplenism.

Authors:  J Zhu; H Zhu; Z Mei; L Zhang; C Jin; L Ran; K Zhou; W Yang
Journal:  Br J Radiol       Date:  2014-08-20       Impact factor: 3.039

4.  Capacitive micromachined ultrasonic transducers for therapeutic ultrasound applications.

Authors:  Serena H Wong; Mario Kupnik; Ronald D Watkins; Kim Butts-Pauly; Butrus T Pierre Khuri-Yakub
Journal:  IEEE Trans Biomed Eng       Date:  2009-07-21       Impact factor: 4.538

5.  Development of a Novel Guided Wave Generation System Using a Giant Magnetostrictive Actuator for Nondestructive Evaluation.

Authors:  Mingzhang Luo; Weijie Li; Junming Wang; Ning Wang; Xuemin Chen; Gangbing Song
Journal:  Sensors (Basel)       Date:  2018-03-04       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.