Literature DB >> 11367769

[Determination of valproic acid level in serum by high performance liquid chromatography].

H Xiao1, S Zhang.   

Abstract

A method for the determination of valproic acid level in serum with high performance liquid chromatography was established. The conditions were Zorbax ODS (250 mm x 4.6 mm, 5 microns) column, 0.8 mL/min sulfuric acid solution (pH 4.2) eluent and UV-210 nm detector. The linear range was 0.82 mumol/L to 1.8 mumol/L (r = 0.996) and the detection limit was 0.82 mumol/L. The average recovery was 98.32% +/- 2.27% and the coefficients of variation were 3.31% (n = 5) and 4.82% (n = 7) for within-day and day-to-day tests respectively. This direct method was compared with the derivatization method.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 11367769

Source DB:  PubMed          Journal:  Se Pu        ISSN: 1000-8713


  1 in total

1.  Modern methods for analysis of antiepileptic drugs in the biological fluids for pharmacokinetics, bioequivalence and therapeutic drug monitoring.

Authors:  Juseop Kang; Yoo-Sin Park; Shin-Hee Kim; Sang-Hyun Kim; Min-Young Jun
Journal:  Korean J Physiol Pharmacol       Date:  2011-04-30       Impact factor: 2.016