| Literature DB >> 11360038 |
T Someya1, K Shimoda, T Muratake, E Nakajima, A Shirai, H Funaoka, T Yashiki, H Ishii, N Sunahara, S Takahashi.
Abstract
Measurement of serum sultopride levels was performed using an enzyme immunoassay. Little or no cross-reactivity with metabolites of sultopride and other drugs was found. The results of reproducibility, recovery, and dilution testing were all good enough for clinical use. A comparison between the measurement values of this method (y) with that of high-performance liquid chromatography (x) showed high correlation (n = 211, r = 0.991, p < 0.0001, y = 0.99x + 107.5). In a comparison between the sultopride dose and serum levels in 161 patients, interindividual differences were large (19 times for same doses), implying that the serum level cannot be predicted from the dosage. The method was found to be reliable for serum level measurements of sultopride and useful for monitoring compliance and assessing the optimal dose.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11360038 DOI: 10.1097/00007691-200106000-00016
Source DB: PubMed Journal: Ther Drug Monit ISSN: 0163-4356 Impact factor: 3.681