Accurate diagnosis of hip prosthesis loosening using a vibrational technique.

A P Georgiou1, J L Cunningham.   

Abstract

OBJECTIVE: To examine the potential role of vibration testing as a non-invasive method of diagnosing loosening of total hip replacements in the clinical setting.
DESIGN: Single blind cohort study in two hospitals.
BACKGROUND: Diagnosing loosening of total hip replacements is heavily dependent on investigative techniques that are unreliable. Previous studies into the use of vibration testing have produced conflicting results.
METHODS: Comparison of vibration testing and radiographs in patients with a total hip replacement experiencing hip pain symptomatic of loosening, with patients showing evidence of a secure prosthesis.
RESULTS: Vibration testing has a sensitivity of 80% and a specificity of 89%. The positive predictive value was 92% and the negative predictive value was 73%; it was unable to produce a definitive diagnosis in 8% of patients. When compared with radiographs from the same patients, vibration testing was shown to be 20% more sensitive and able to diagnose 13% more patients.
CONCLUSIONS: Vibration testing can deliver more accurate information on the stability of total hip replacements than radiographs in the clinical setting, despite being in the early stages of development. Relevance. This study shows that 70 more patients may be provided with the correct diagnosis and 46 fewer patients may be undiagnosed each year, when using vibrometry as opposed to radiographs. In view of the relative disparity between the level of development between the two techniques and the encouraging results hitherto presented, it is felt that by improving vibration testing it may supersede radiographs in the detection of prosthesis loosening.

Entities:  

Mesh:

Year:  2001        PMID: 11358619     DOI: 10.1016/s0268-0033(01)00002-x

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  10 in total

1.  Knee Implant Loosening Detection: A Vibration Analysis Investigation.

Authors:  Arash Arami; Jean-Romain Delaloye; Hossein Rouhani; Brigitte M Jolles; Kamiar Aminian
Journal:  Ann Biomed Eng       Date:  2017-10-24       Impact factor: 3.934

2.  A study on the use of the Osstell apparatus to evaluate pedicle screw stability: An in-vitro study using micro-CT.

Authors:  Daisuke Nakashima; Ken Ishii; Morio Matsumoto; Masaya Nakamura; Takeo Nagura
Journal:  PLoS One       Date:  2018-06-28       Impact factor: 3.240

3.  Design, Modeling, and Evaluation of the Eddy Current Sensor Deeply Implanted in the Human Body.

Authors:  Rajas Prakash Khokle; Karu P Esselle; Desmond J Bokor
Journal:  Sensors (Basel)       Date:  2018-11-11       Impact factor: 3.576

Review 4.  Altering the Course of Technologies to Monitor Loosening States of Endoprosthetic Implants.

Authors:  João Henrique Cachão; Marco P Soares Dos Santos; Rodrigo Bernardo; António Ramos; Rainer Bader; Jorge A F Ferreira; António Torres Marques; José A O Simões
Journal:  Sensors (Basel)       Date:  2019-12-23       Impact factor: 3.576

5.  Multiscale Sensing of Bone-Implant Loosening for Multifunctional Smart Bone Implants: Using Capacitive Technologies for Precision Controllability.

Authors:  Inês Peres; Pedro Rolo; Jorge A F Ferreira; Susana C Pinto; Paula A A P Marques; António Ramos; Marco P Soares Dos Santos
Journal:  Sensors (Basel)       Date:  2022-03-25       Impact factor: 3.576

6.  Investigation of a passive sensor array for diagnosis of loosening of endoprosthetic implants.

Authors:  Cathérine Ruther; Christian Schulze; Andrea Boehme; Hannes Nierath; Hartmut Ewald; Wolfram Mittelmeier; Rainer Bader; Daniel Kluess
Journal:  Sensors (Basel)       Date:  2012-12-20       Impact factor: 3.576

7.  Acoustic monitoring (RFM) of total hip arthroplasty - Results of a cadaver study.

Authors:  A C Unger; H Cabrera-Palacios; A P Schulz; Ch Jürgens; Andreas Paech
Journal:  Eur J Med Res       Date:  2009-06-18       Impact factor: 2.175

8.  In vivo evaluation of a vibration analysis technique for the per-operative monitoring of the fixation of hip prostheses.

Authors:  Leonard C Pastrav; Siegfried Vn Jaecques; Ilse Jonkers; Georges Van der Perre; Michiel Mulier
Journal:  J Orthop Surg Res       Date:  2009-04-09       Impact factor: 2.359

9.  Laser Resonance Frequency Analysis: A Novel Measurement Approach to Evaluate Acetabular Cup Stability During Surgery.

Authors:  Shunsuke Kikuchi; Katsuhiro Mikami; Daisuke Nakashima; Toshiyuki Kitamura; Noboru Hasegawa; Masaharu Nishikino; Arihiko Kanaji; Masaya Nakamura; Takeo Nagura
Journal:  Sensors (Basel)       Date:  2019-11-08       Impact factor: 3.576

10.  Laser resonance frequency analysis of pedicle screw stability: A cadaveric model bone study.

Authors:  Daisuke Nakashima; Katsuhiro Mikami; Shunsuke Kikuchi; Masaharu Nishikino; Toshiyuki Kitamura; Noboru Hasegawa; Morio Matsumoto; Masaya Nakamura; Takeo Nagura
Journal:  J Orthop Res       Date:  2021-01-28       Impact factor: 3.494