Different strategies for Humphrey automated perimetry: FASTPAC, SITA standard and SITA fast in normal subjects and glaucoma patients.

X Roggen1, K Herman, L Van Malderen, M Devos, W Spileers.   

Abstract

PURPOSE: To evaluate the influence on examination time and test quality of the recently introduced SITA strategies of the Humphrey Field Analyzer.
METHODS: The sample consisted of 41 subjects (19 normal subjects and 22 glaucoma patients), all experienced in automated perimetry, ranging in age from 24 to 83 years. One eye of each patient was examined with the HC30-2 program using the FASTPAC (FP) and SITA Standard (SS) or SITA Fast (SF) strategy on the same day, in random order. Examination time was evaluated as a function of the strategy. To evaluate the test quality both regional and global visual field parameters were analyzed. Global parameters included mean deviation (MD) and pattern standard deviation (PSD). Regional parameters (mean and maximum loss) were calculated to estimate the extent and the depth of localized visual field defects. For this purpose each visual field was divided in 4 quadrants and in 10 clusters as defined in the glaucoma hemifield test.
RESULTS: 1. There is a considerable test time reduction from FASTPAC over SITA Standard to SITA Fast for comparable MD and PSD values. On average, the SITA Fast test duration is half that of the FASTPAC procedure. For each strategy, the test duration increases for increasing visual field loss. 2. Between all three strategies, there is a good correlation for the global indices (MD, PSD) 3. For the regional indices (mean loss, maximum loss) the same high correlation exists.
CONCLUSION: The SITA strategy causes a significant test time reduction without decreasing the test quality.

Entities:  

Mesh:

Year:  2001        PMID: 11344712

Source DB:  PubMed          Journal:  Bull Soc Belge Ophtalmol        ISSN: 0081-0746


  3 in total

1.  Swedish Interactive Threshold Algorithm for central visual field defects unrelated to nerve fiber layer.

Authors:  Kazunori Hirasawa; Nobuyuki Shoji
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-08-18       Impact factor: 3.117

2.  Comparison of optic nerve head topography findings in eyes with non-arteritic anterior ischemic optic neuropathy and eyes with glaucoma.

Authors:  Josepha Horowitz; Tagil Fishelzon-Arev; Eitan Z Rath; Eitan Segev; Orna Geyer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-03-06       Impact factor: 3.117

Review 3.  Automated achromatic perimetry.

Authors:  Anand Aggarwal; Kanika Chhabra; Prempal Kaur; Karamjit Singh; Indu Khosa; Pulkit Bansal
Journal:  Oman J Ophthalmol       Date:  2018 Jan-Apr
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.