Literature DB >> 11341083

Incidence of electromagnetic interference in implantable cardioverter defibrillators.

C Kolb1, B Zrenner, C Schmitt.   

Abstract

Electromagnetic interference (EMI) with ICDs can lead to temporary inhibition of the device or to inappropriate delivery of antitachycardia pacing and shocks. The incidence of interactions between electronic devices and the current generation of ICDs is not known. In a retrospective study of 341 patients (665 patient-years) who underwent a regular follow-up every 3 months, five episodes of EMI were detected in four different patients. The risk for receiving inappropriate shocks due to EMI is < 1% per year and patient. In conclusion, although inappropriate delivery of shocks by ICDs due to EMI rarely occurs, patient information should emphasize the avoidance of situations of possible interference. Further efforts concerning lead technology and detection algorithms are necessary to minimize the risk of EMI.

Entities:  

Mesh:

Year:  2001        PMID: 11341083     DOI: 10.1046/j.1460-9592.2001.00465.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  11 in total

1.  Inappropriate shock delivery due to interference between a washing machine and an implantable cardioverter defibrillator.

Authors:  Christof Kolb; Sebastian Schmieder; Claus Schmitt
Journal:  J Interv Card Electrophysiol       Date:  2002-12       Impact factor: 1.900

2.  Incidence of magnet mode in patients with implantable cardioverter defibrillators.

Authors:  Gesa von Olshausen; Johanna Schorr; Christian Grebmer; Severin Weigand; Patrick Blažek; Amir Brkic; Lena Friedrich; Verena Semmler; Christof Kolb; Carsten Lennerz
Journal:  J Interv Card Electrophysiol       Date:  2019-10-16       Impact factor: 1.900

Review 3.  Troubleshooting implantable cardioverter-defibrillators: an overview for physicians who are not electrophysiologists.

Authors:  Mohammad Saeed
Journal:  Tex Heart Inst J       Date:  2011

4.  Safe scanning, but frequent artifacts mimicking bradycardia and tachycardia during magnetic resonance imaging (MRI) in patients with an implantable loop recorder (ILR).

Authors:  J Rod Gimbel; Jamal Zarghami; Christian Machado; Bruce L Wilkoff
Journal:  Ann Noninvasive Electrocardiol       Date:  2005-10       Impact factor: 1.468

5.  Inappropriate ICD discharge induced by electrical interference from a physio-therapeutic muscle stimulation device.

Authors:  H Nägele; M Azizi
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2006-09

6.  Correlation of geomagnetic activity with implantable cardioverter defibrillator shocks and antitachycardia pacing.

Authors:  Elisa Ebrille; Tomas Konecny; Dana Konecny; Radim Spacek; Paul Jones; Pavel Ambroz; Christopher V DeSimone; Brian D Powell; David L Hayes; Paul A Friedman; Samuel J Asirvatham
Journal:  Mayo Clin Proc       Date:  2015-02       Impact factor: 7.616

Review 7.  [Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].

Authors:  Bernd Nowak; Oliver Przibille; Andreas Napp
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2022-07-04

8.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

9.  Electromagnetic interference in implantable cardioverter defibrillators: present but rare.

Authors:  Gesa von Olshausen; Ina-Christine Rondak; Carsten Lennerz; Verena Semmler; Christian Grebmer; Tilko Reents; Sonia Ammar-Busch; Alessandra Buiatti; Felix Bourier; Isabel Deisenhofer; Christof Kolb
Journal:  Clin Res Cardiol       Date:  2016-01-29       Impact factor: 5.460

10.  Frequent Home Monitoring of ICD Is Effective to Prevent Inappropriate Defibrillator Shock Delivery.

Authors:  Paolo Bifulco; Luigi Argenziano; Maria Romano; Mario Cesarelli; Mario Sansone; Stefano Casella; Stefano Nardi
Journal:  Case Rep Med       Date:  2014-01-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.