In vitro hemodynamic evaluation of a Simon nitinol vena cava filter: possible explanation of IVC occlusion.

R L Leask1, K W Johnston, M Ojha.   

Abstract

PURPOSE: To evaluate the local hemodynamics in the region of the Simon nitinol filter (SNF), used to prevent pulmonary emboli by capturing clot and promoting lysis.
MATERIALS AND METHODS: The hemodynamics of the Simon nitinol inferior vena cava (IVC) filter were evaluated under steady flow (Re = 600) in a 20-mm-diameter IVC model. The photochromic dye tracer technique was used to estimate the velocity and wall shear stress. These flow features were determined for the unoccluded and partially occluded (clot volume = 1,500 mm(3)) states of the SNF along its center plane.
RESULTS: A region of low velocities developed around the central axis of the filter extending from the leading edge of the central strut to the filter tip. This phenomenon was created by the strong redirection of flow toward the periphery of the filter. With the presence of the clot, these effects were enhanced, causing flow separation and recirculation. In addition, the shear stress on the hip of the clot was about 30 times that of the upstream value, and turbulence developed in the near-downstream region.
CONCLUSIONS: The extended region of almost-stagnant flow near the midsection of the umbrella region could lead to organization of thrombus and fibrin mesh network development. The presence of a simulated clot led to a significant increase in the size of the stagnant, thrombus-prone region as well as turbulence, which, overall, may contribute to caval occlusion.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11340141     DOI: 10.1016/s1051-0443(07)61486-9

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  2 in total

1.  Modeling hemodynamics in an unoccluded and partially occluded inferior vena cava under rest and exercise conditions.

Authors:  Zhuyin Ren; Stephen L Wang; Michael A Singer
Journal:  Med Biol Eng Comput       Date:  2012-02-22       Impact factor: 2.602

2.  Hemodynamic Analysis of VenaTech Convertible Vena Cava Filter Using Computational Fluid Dynamics.

Authors:  Jingying Wang; Wen Huang; Yue Zhou; Fangzhou Han; Dong Ke; Chunhian Lee
Journal:  Front Bioeng Biotechnol       Date:  2020-10-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.