V Bucher1, B Brunner, C Leibrock, M Schubert, W Nisch.
Abstract
A light-addressable microelectrode chip with 3600 TiN electrodes was fabricated. Amorphous silicon (a-Si:H) serves as a photo conductor. The electrodes on the chip are addressed by a laser spot and electrical properties of the system are determined. DC measurements show a dark to bright dynamic of 10(6)-10(7). The AC impedance dynamic @ 1 kHz/100 mV and thus the signal-to-noise-ratio is determined to 60. This value is quite sufficient for electrophysiological measurements. For the first time, recordings from cardiac myocytes are reported using the principle of light-addressing. Measurements were done with a standard laser scan microscope (Zeiss LSM 410).Entities:
Mesh:
Substances:
Year: 2001 PMID: 11339999 DOI: 10.1016/s0956-5663(01)00135-x