Reproducibility of ductus venosus Doppler flow measurements at 11-14 weeks of gestation.

F Prefumo1, P De Biasio, P L Venturini.   

Abstract

OBJECTIVES: To assess the intra- and interobserver repeatabilities of fetal ductus venosus Doppler measurements at 11-14 weeks of gestation.
DESIGN: Flow velocity waveforms were recorded transabdominally. Intraobserver repeatability was studied in 22 fetuses in whom four repeated measurements were performed by the same observer. Interobserver repeatability was assessed in 54 fetuses in each of whom two observers performed two repeated measurements. The pulsatility index for veins (PIV), peak velocity during ventricular systole (S wave), peak velocity during atrial contraction (A wave) and time-averaged maximum velocity (TAMV) were recorded. Reproducibility of the Doppler measurements was analyzed by calculating repeatability coefficient, coefficient of variation (CV), intraclass correlation coefficient (ICC), mean differences and their limits of agreement. Cohen's k-coefficient was used for categorical data.
RESULTS: Intraobserver repeatability was good with a CV of 10%, 13%, 22% and 13% and an ICC of 0.85, 0.94, 0.94 and 0.95, respectively, for PIV, S wave, A wave and TAMV. Interobserver repeatability was also good, and no bias between the observers was detected. A CV of 8.8%, 14%, 27% and 15% and an ICC of 0.86, 0.84, 0.87 and 0.84, respectively, for PIV, S wave, A wave and TAMV were obtained. There was 100% agreement for the detection of normal/abnormal blood flow.
CONCLUSIONS: Intra- and interobserver repeatability of all parameters was acceptable, allowing for the detection of moderate to large changes in Doppler measurements.

Entities:  

Mesh:

Year:  2001        PMID: 11339185     DOI: 10.1046/j.1469-0705.2001.00338.x

Source DB:  PubMed          Journal:  Ultrasound Obstet Gynecol        ISSN: 0960-7692            Impact factor:   7.299


  2 in total

1.  Doppler assessment of the ductus venosus and the tricuspid valve at 11-13+6 weeks: Reference ranges and development of sonographic quality assurance standards.

Authors:  Vanessa Pincham; Jon Hyett; Karen Pollard; Philip Schluter; Andrew McLennan
Journal:  Australas J Ultrasound Med       Date:  2016-02-21

2.  Reproducibility and repeatability of measuring the electrical impedance of the pregnant human cervix-the effect of probe size and applied pressure.

Authors:  Roobin P Jokhi; Vidita V Ghule; Brian H Brown; Dilly O C Anumba
Journal:  Biomed Eng Online       Date:  2009-06-17       Impact factor: 2.819

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.