Quantitative assessment of motion artifacts and validation of a new motion-correction program for myocardial perfusion SPECT.

N Matsumoto1, D S Berman, P B Kavanagh, J Gerlach, S W Hayes, H C Lewin, J D Friedman, G Germano.   

Abstract

UNLABELLED: Patient motion during myocardial perfusion SPECT can produce images that show artifactual perfusion defects. The relationship between the degree of motion and the extent of artifactual perfusion defects is not clear for either single- or double-head detectors. Using both single- and double-head detectors and quantitative perfusion SPECT (QPS) software, we studied the pattern and extent of defects induced by simulated motion and validated a new automatic motion-correction program for myocardial perfusion SPECT.
METHODS: Vertical motion was simulated by upward shifting of the raw projection datasets in a returning pattern (bounce) and in a nonreturning pattern at 3 different phases of the SPECT acquisition (early, middle, and late), whereas upward creep was simulated by uniform shifting throughout the acquisition. Lateral motion was similarly simulated by left shifting of the raw projection datasets in a returning pattern and in a nonreturning pattern. Simulations were performed using single- and double-head detectors, and simulated motion was applied to projection images from 8 patients who had normal 99mTc-sestamibi SPECT findings. Additionally, images from 130 patients with actual clinical motion were assessed before and after motion correction. The extent of perfusion defects was assessed by QPS, and a 20-segment, 5-point scoring system was used to assess the effect of motion on the presence and extent of perfusion defects.
RESULTS: Of 12 bounce simulations, the bouncing motion failed to produce significant (>3%) perfusion defects with either the single- or the double-head detector. With the single-head detector, early shifting created the largest defect, whereas with the double-head detector, shifting during the middle of the acquisition created the largest defect. With regard to upward creep, defects were of larger extent with the double- than the single-head detector. With the single-head detector, 8 of 20 simulated motion patterns yielded significant perfusion defects of the left ventricle, 7 (88%) of which were significantly improved after motion correction. With the double-head detector, 12 of 20 patterns yielded significant defects, all of which improved significantly after correction. Of 2,600 segments in the 130 patients with actual clinical motion, only 1.3% (30/2,259) of segments that were considered normal (score = 0 or 1) changed to abnormal (score = 2-4) after motion correction, whereas 27% (92/341) of abnormal segments were reclassified as normal after motion correction.
CONCLUSION: Artifactual perfusion defects created by simulated motion are a function of the time, degree, and type of motion and the number of camera detectors. Application of an automatic motion-correction algorithm effectively decreases motion artifacts on myocardial perfusion SPECT images.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11337561

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  41 in total

Review 1.  Effect of motion on cardiac SPECT imaging: recognition and motion correction.

Authors:  J Fitzgerald; P G Danias
Journal:  J Nucl Cardiol       Date:  2001 Nov-Dec       Impact factor: 5.952

Review 2.  Quantitative analysis of perfusion studies: strengths and pitfalls.

Authors:  Piotr Slomka; Yuan Xu; Daniel Berman; Guido Germano
Journal:  J Nucl Cardiol       Date:  2012-04       Impact factor: 5.952

3.  A study to quantify the effect of patient motion and develop methods to detect and correct for motion during myocardial perfusion imaging on a CZT solid-state dedicated cardiac camera.

Authors:  Shelley Redgate; David C Barber; John W Fenner; Abdallah Al-Mohammad; Jonathon C Taylor; Michael B Hanney; Wendy B Tindale
Journal:  J Nucl Cardiol       Date:  2015-12-18       Impact factor: 5.952

4.  EANM/ESC procedural guidelines for myocardial perfusion imaging in nuclear cardiology.

Authors:  B Hesse; K Tägil; A Cuocolo; C Anagnostopoulos; M Bardiés; J Bax; F Bengel; E Busemann Sokole; G Davies; M Dondi; L Edenbrandt; P Franken; A Kjaer; J Knuuti; M Lassmann; M Ljungberg; C Marcassa; P Y Marie; F McKiddie; M O'Connor; E Prvulovich; R Underwood; B van Eck-Smit
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-07       Impact factor: 9.236

5.  Use of three-dimensional Gaussian interpolation in the projector/backprojector pair of iterative reconstruction for compensation of known rigid-body motion in SPECT.

Authors:  Bing Feng; Howard C Gifford; Richard D Beach; Guido Boening; Michael A Gennert; Michael A King
Journal:  IEEE Trans Med Imaging       Date:  2006-07       Impact factor: 10.048

6.  Efficacy of emergent transcatheter transplantation of stem cells for treatment of acute myocardial infarction (TCT-STAMI).

Authors:  J Ge; Y Li; J Qian; J Shi; Q Wang; Y Niu; B Fan; X Liu; S Zhang; A Sun; Y Zou
Journal:  Heart       Date:  2006-06-14       Impact factor: 5.994

Review 7.  Optimal SPECT processing and display: making bad studies look good to get the right answer.

Authors:  Dalia Y Ibrahim; Frank P DiFilippo; Jeremy E Steed; Manuel D Cerqueira
Journal:  J Nucl Cardiol       Date:  2006-11       Impact factor: 5.952

Review 8.  Gated SPECT in assessment of regional and global left ventricular function: major tool of modern nuclear imaging.

Authors:  Aiden Abidov; Guido Germano; Rory Hachamovitch; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

9.  Prediction of revascularization after myocardial perfusion SPECT by machine learning in a large population.

Authors:  Reza Arsanjani; Damini Dey; Tigran Khachatryan; Aryeh Shalev; Sean W Hayes; Mathews Fish; Rine Nakanishi; Guido Germano; Daniel S Berman; Piotr Slomka
Journal:  J Nucl Cardiol       Date:  2014-12-06       Impact factor: 5.952

10.  Two-position supine/prone myocardial perfusion SPECT (MPS) imaging improves visual inter-observer correlation and agreement.

Authors:  Reza Arsanjani; Sean W Hayes; Mathews Fish; Aryeh Shalev; Rine Nakanishi; Louise E J Thomson; John D Friedman; Guido Germano; Daniel S Berman; Piotr Slomka
Journal:  J Nucl Cardiol       Date:  2014-05-08       Impact factor: 5.952

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.