Performance of the VITEK2 system for identification and susceptibility testing of routine Enterobacteriaceae clinical isolates.

M Pérez-Vázquez1, A Oliver, B Sánchez del Saz, E Loza, F Baquero, R Cantón.   

Abstract

The VITEK2 system was evaluated with 138 fresh consecutive routine clinical Enterobacteriaceae isolates. Susceptibility results to 10 beta-lactams, three aminoglycosides and a quinolone were compared with those obtained following the NCCLS standard microdilution. API20E was used as reference method for identification. All but three isolates were correctly identified in 3 h at species level (97.8%), two isolates (1.4%) at genus level and only one isolate was misidentified. Overall essential agreement for susceptibility testing was 97.1%. Discrepancies were mainly observed with piperacillin (1.1%), cefuroxime (0.6%) and amoxycillin/clavulanate (0.3%). Discrepancies for aminoglycosides and ciprofloxacin were low (<0.1%). Minor, major and very major errors (NCCLS categories) were 4.1%, 0.2% and 6.1%, respectively. Very major errors were due to piperacillin (4.5%), ampicillin (0.8%) and amoxycillin/clavulanate (0.8%). The VITEK2 system gave accurate identification and susceptibility testing results of routine Enterobacteriaceae clinical isolates.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11337223     DOI: 10.1016/s0924-8579(01)00318-1

Source DB:  PubMed          Journal:  Int J Antimicrob Agents        ISSN: 0924-8579            Impact factor:   5.283


  7 in total

1.  Analysis of the comparative workflow and performance characteristics of the VITEK 2 and Phoenix systems.

Authors:  U Eigner; A Schmid; U Wild; D Bertsch; A-M Fahr
Journal:  J Clin Microbiol       Date:  2005-08       Impact factor: 5.948

Review 2.  Expert systems in clinical microbiology.

Authors:  Trevor Winstanley; Patrice Courvalin
Journal:  Clin Microbiol Rev       Date:  2011-07       Impact factor: 26.132

3.  Performance evaluation of the Verigene® (Nanosphere) and FilmArray® (BioFire®) molecular assays for identification of causative organisms in bacterial bloodstream infections.

Authors:  C Ward; K Stocker; J Begum; P Wade; U Ebrahimsa; S D Goldenberg
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2014-10-14       Impact factor: 3.267

4.  The Poisoned Well: Enhancing the Predictive Value of Antimicrobial Susceptibility Testing in the Era of Multidrug Resistance.

Authors:  Thea Brennan-Krohn; Kenneth P Smith; James E Kirby
Journal:  J Clin Microbiol       Date:  2017-05-03       Impact factor: 5.948

5.  Septic arthritis of the knee due to Raoultella ornithinolytica.

Authors:  Kevin Venus; Sarah Vaithilingam; Isaac I Bogoch
Journal:  Infection       Date:  2016-08-03       Impact factor: 3.553

6.  Rapid detection of methicillin resistance in coagulase-negative Staphylococci with the VITEK 2 system.

Authors:  Matthias A Horstkotte; Johannes K-M Knobloch; Holger Rohde; Sabine Dobinsky; Dietrich Mack
Journal:  J Clin Microbiol       Date:  2002-09       Impact factor: 5.948

7.  Quality of water and antibiotic resistance of Escherichia coli from water sources of hilly tribal villages with and without integrated watershed management-a one year prospective study.

Authors:  Sandeep S Nerkar; Ashok J Tamhankar; Smita U Khedkar; Cecilia Stålsby Lundborg
Journal:  Int J Environ Res Public Health       Date:  2014-06       Impact factor: 3.390

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.