| Literature DB >> 11334355 |
P de Bruijn1, D F Kehrer, J Verweij, A Sparreboom.
Abstract
A high-performance liquid chromatographic assay with UV detection has been developed for the determination of ketoconazole in human plasma. Quantitative extraction was achieved by a single solvent extraction involving a mixture of acetonitrile-n-butyl chloride (1:4, v/v). Ketoconazole and the internal standard (clotrimazole) were separated on a column packed with Inertsil ODS-80A material and a mobile phase composed of water-acetonitrile-tetrahydrofuran-ammonium hydroxide-triethylamine (45:50.2:2.5:0.1:0.1, v/v). The column effluent was monitored at a wavelength of 206 nm with a detector range set at 0.5. The calibration graph was linear in the range of 20-2000 ng/ml, with a lower limit of quantitation of 20.0 ng/ml. The extraction recoveries for ketoconazole and clotrimazole in human plasma were 93+/-9.7% and 83+/-10.0%, respectively. The developed method has been successfully applied to a clinical study to examine the pharmacokinetics of ketoconazole in a cancer patient.Entities:
Mesh:
Substances:
Year: 2001 PMID: 11334355 DOI: 10.1016/s0378-4347(00)00573-9
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273