Functional differences in bi-level pressure preset ventilators.

M P Highcock1, J M Shneerson, I E Smith.   

Abstract

The performance of four bilevel positive pressure preset ventilators was compared. The ventilators tested were; BiPAP ST30 (Respironics); Nippy2 (B + D Electrical); Quantum PSV (Healthdyne); and Sullivan VPAP H ST (Resmed). A patient simulator was used to determine the sensitivity of the triggering mechanisms and the responses to a leak within the patient circuit, and to changes in patient effort. Significant differences (p <0.05) between the devices were seen in the trigger delay time and inspiratory trigger pressure. When a leak was introduced into the patient circuit, the fall in tidal volume (VT) was less than ten per cent for each ventilator. The addition of patient effort produced a number of changes in the ventilation delivered. Patient efforts of 0.25 s induced a variable fall in VT. An increase in VT was seen with some ventilators with patient efforts of 1 s but the effect was variable. Trigger failures and subsequent falls in minute volume were seen with the BiPAP and the Nippy2 at the highest respiratory frequency. Differences in the responses of the ventilators are demonstrated that may influence the selection of a ventilator, particularly in the treatment of breathless patients with ventilatory failure.

Entities:  

Mesh:

Year:  2001        PMID: 11334130     DOI: 10.1183/09031936.01.17202680

Source DB:  PubMed          Journal:  Eur Respir J        ISSN: 0903-1936            Impact factor:   16.671


  3 in total

1.  Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives.

Authors:  Carlo Olivieri; Roberta Costa; Giorgio Conti; Paolo Navalesi
Journal:  Intensive Care Med       Date:  2011-11-29       Impact factor: 17.440

2.  The configuration of bi-level ventilator circuits may affect compensation for non-intentional leaks during volume-targeted ventilation.

Authors:  Annalisa Carlucci; Annia Schreiber; Alessio Mattei; Alberto Malovini; Jacopo Bellinati; Piero Ceriana; Cesare Gregoretti
Journal:  Intensive Care Med       Date:  2012-09-28       Impact factor: 17.440

3.  Performance characteristics of seven bilevel mechanical ventilators in pressure-support mode with different cycling criteria: a comparative bench study.

Authors:  Yuqing Chen; Kewen Cheng; Xin Zhou
Journal:  Med Sci Monit       Date:  2015-01-26
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.