Literature DB >> 11328916

Accuracy and safety of online clearance monitoring based on conductivity variation.

U Kuhlmann1, R Goldau, N Samadi, T Graf, M Gross, G Orlandini, H Lange.   

Abstract

BACKGROUND: Haemodialysis dose has been shown to have a distinct impact upon the morbidity and mortality rate in patients on regular dialysis therapy. Accordingly, the adequacy of dialysis treatment should be guaranteed.
METHODS: In 200 dialysis sessions two or three +/-10% dialysate conductivity variations were applied to test patient compliance and the accuracy of an electrolyte based online clearance measurement (OCM) reflecting the total clearance of urea.
RESULTS: Using a step profile the electrolytic clearance showed highly significant correlation with the reference data in the blood side (n=118, r=0.867, P<0.001) and dialysate side (n=118, r=0.820, P<0.001) if only reference values were taken into account for which the error in mass balance did not exceed 5%. Kt/V according to the single pool model (n=35, r=0.940, P<0.001), the equilibrated single pool variable volume kinetic model (n=36, r=0.982, P<0.001), Daugirdas formula (n=34, r=0.951, P<0.001) and direct quantification of dialysance via spent dialysate (n=26, r=0.900, P<0.001) showed outstanding correlations with electrolyte-based Kt/V at mass balance error below 5%. No adverse clinical effect of OCM was reported. Serum sodium, body weight, heart rate and breathing rate at rest, arterial pO(2) and pCO(2) and blood pressure before haemodialysis remained unaffected in OCM measurements in comparison with baseline parameters. A small influx of sodium (1.53+/-7.62 mmol) into the patient was seen following the impulse, but no signs associated with fluid overload were observed during the study period of 10 consecutive dialysis sessions.
CONCLUSIONS: The OCM option of the haemodialysis machine provides a safe and accurate tool for continuous online monitoring of total urea clearance.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11328916     DOI: 10.1093/ndt/16.5.1053

Source DB:  PubMed          Journal:  Nephrol Dial Transplant        ISSN: 0931-0509            Impact factor:   5.992


  11 in total

1.  Ionic dialysance: a new valid parameter for quantification of dialysis efficiency in acute renal failure?

Authors:  Christophe Ridel; David Osman; Lucile Mercadal; Nadia Anguel; Thierry Petitclerc; Christian Richard; Christophe Vinsonneau
Journal:  Intensive Care Med       Date:  2007-01-18       Impact factor: 17.440

2.  Oxalate quantification in hemodialysate to assess dialysis adequacy for primary hyperoxaluria.

Authors:  Xiaojing Tang; Nikolay V Voskoboev; Stacie L Wannarka; Julie B Olson; Dawn S Milliner; John C Lieske
Journal:  Am J Nephrol       Date:  2014-04-26       Impact factor: 3.754

3.  Prospective Study of Routine Heparin Avoidance Hemodialysis in a Tertiary Acute Care Inpatient Practice.

Authors:  Sami Safadi; Robert C Albright; John J Dillon; Amy W Williams; Fares Alahdab; Julie K Brown; Amanda L Severson; Walter K Kremers; Mary Ann Ryan; Marie C Hogan
Journal:  Kidney Int Rep       Date:  2017-03-16

4.  Interventions to improve hemodialysis adequacy: protocols based on real-time monitoring of dialysate solute clearance.

Authors:  Edward A Ross; Jennifer L Paugh-Miller; Robert W Nappo
Journal:  Clin Kidney J       Date:  2017-10-25

5.  Monitoring dialysis adequacy: history and current practice.

Authors:  Linda Ding; James Johnston; Maury N Pinsk
Journal:  Pediatr Nephrol       Date:  2021-01-05       Impact factor: 3.714

6.  The Kt/V by ionic dialysance: Interpretation limits.

Authors:  A Alayoud; D Montassir; A Hamzi; Y Zajjari; A Bahadi; D El Kabbaj; O Maoujoud; T Aatif; K Hassani; M Benyahia; Z Oualim
Journal:  Indian J Nephrol       Date:  2012-09

7.  The effect of citrate dialysate on intradialytic heparin dose in haemodialysis patients: study design of a randomised controlled trial.

Authors:  Davina J Tai; Kelvin Leung; Pietro Ravani; Robert R Quinn; Nairne Scott-Douglas; Jennifer M MacRae
Journal:  BMC Nephrol       Date:  2015-08-25       Impact factor: 2.388

8.  Ultrafiltration biofeedback guided by blood volume monitoring to reduce intradialytic hypotensive episodes in hemodialysis: study protocol for a randomized controlled trial.

Authors:  Kelvin C W Leung; Robert R Quinn; Pietro Ravani; Jennifer M MacRae
Journal:  Trials       Date:  2014-12-10       Impact factor: 2.279

9.  Functional training added to intradialytic cycling lowers low-density lipoprotein cholesterol and improves dialysis adequacy: a randomized controlled trial.

Authors:  Špela Bogataj; Jernej Pajek; Jadranka Buturović Ponikvar; Maja Pajek
Journal:  BMC Nephrol       Date:  2020-08-18       Impact factor: 2.388

10.  Correlation between Dt/V derived from ionic dialysance and blood-driven Kt/V of urea in African-American hemodialysis patients, based on body weight and ultrafiltration volume.

Authors:  Wihib Gebregeorgis; Zeenat Yousuf Bhat; Nishigandha Pradhan; Stephen D Migdal; Lakshminarayanan Nandagopal; Reddy Singasani; Tehmina Mushtaq; Ronald Thomas; Yahya M Osman Malik
Journal:  Clin Kidney J       Date:  2018-01-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.