Detection and quantification of intraperitoneal fluid using electrical impedance tomography.

R J Sadleir1, R A Fox.   

Abstract

A prototype electrical impedance tomography system was evaluated prior to its use for the detection of intraperitoneal bleeding, with the assistance of patients undergoing continuous ambulatory peritoneal dialysis (CAPD). The system was sensitive enough to detect small amounts of dialysis fluid appearing in subtractive images over short time periods. Uniform sensitivity to blood appearing anywhere within the abdominal cavity was produced using a post-reconstructive filter that corrected for changes in apparent resistivity of anomalies with their radial position. The image parameter used as an indication of fluid quantity, the resistivity index, varied approximately linearly with the quantity of fluid added. A test of the system's response to the introduction of conductive fluid out of the electrode plane (when a blood-equivalent fluid was added to the stomach) found that the sensitivity of the system was about half that observed in the electrode plane. Breathing artifacts were found to upset quantitative monitoring of intraperitoneal bleeding, but only on time scales short compared with the fluid administration rate. Longer term breathing changes, such as those due to variations in the functional residual capacity of the lungs, should ultimately limit the sensitivity over long time periods.

Entities:  

Mesh:

Year:  2001        PMID: 11322536     DOI: 10.1109/10.915715

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  5 in total

1.  Use of 3-D magnetic resonance electrical impedance tomography in detecting human cerebral stroke: a simulation study.

Authors:  Nuo Gao; Shan-An Zhu; Bin He
Journal:  J Zhejiang Univ Sci B       Date:  2005-05       Impact factor: 3.066

2.  A bio-impedance quantitative method based on magnetic induction tomography for intracranial hematoma.

Authors:  Li Ke; Wanni Zu; Qiang Du; Jia Chen; Xiaodi Ding
Journal:  Med Biol Eng Comput       Date:  2020-02-15       Impact factor: 2.602

3.  Use of electrical impedance tomography to monitor regional cerebral edema during clinical dehydration treatment.

Authors:  Feng Fu; Bing Li; Meng Dai; Shi-Jie Hu; Xia Li; Can-Hua Xu; Bing Wang; Bin Yang; Meng-Xing Tang; Xiu-Zhen Dong; Zhou Fei; Xue-Tao Shi
Journal:  PLoS One       Date:  2014-12-04       Impact factor: 3.240

4.  Intravenous administration of normal saline may be misinterpreted as a change of end-expiratory lung volume when using electrical impedance tomography.

Authors:  Vladimír Sobota; Martin Müller; Karel Roubík
Journal:  Sci Rep       Date:  2019-04-08       Impact factor: 4.379

5.  In vivo imaging of twist drill drainage for subdural hematoma: a clinical feasibility study on electrical impedance tomography for measuring intracranial bleeding in humans.

Authors:  Meng Dai; Bing Li; Shijie Hu; Canhua Xu; Bin Yang; Jianbo Li; Feng Fu; Zhou Fei; Xiuzhen Dong
Journal:  PLoS One       Date:  2013-01-25       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.