Open-heart endocardial radiofrequency ablation: an alternative to incisions in Maze surgery.

J A Caccitolo1, J M Stulak, H V Schaff, D Francischelli, D N Jensen, R Mehra.   

Abstract

BACKGROUND: Radiofrequency (RF) ablation produces transmural atrial lesions in vitro, and may provide advantages over incisions currently used in maze surgery. This study examines the feasibility, safety, and efficacy of open-heart endocardial RF ablation.
METHODS: Eighteen sheep (42.8 +/- 4.4 kg, age < 2 years) underwent left thoracotomy with placement of pacing leads on a pulmonary vein and the left atrial dome. On cardiopulmonary bypass, lesions were made using incision and suture or a novel RF ablation device in three sites: PVC = circle excluding pulmonary veins, IAB = line across the interatrial bundle, SVC = line from the superior to the inferior vena cava. Pacing across the PVC lesion was attempted to assess the completeness of each lesion. Preselected animals (incision n = 4, RF n = 5) were recovered and pacing attempts were repeated at 1 month. After sacrifice, hearts were sectioned and measured for lesion size and completeness.
RESULTS: RF ablation lesions took less time to create (total bypass time: RF 51.8 min vs incision 106 min, P < 0.001). No evidence of thromboembolism, atrial rupture, or coronary sinus thrombosis was seen. All PVC lesions were complete as demonstrated by the inability to pace across them. Stained sections demonstrated that acutely studied incision lesions were thinner than RF lesions; however, all lesions were transmural and similar in width at 1 month.
CONCLUSIONS: RF ablation consistently created transmural lesions more quickly than the incision and suture method and without additional complications. Endocardial RF ablation appears to be a simple and effective alternative to surgical incisions during open-heart atrial Maze procedures. Copyright 2001 Academic Press.

Entities:  

Mesh:

Year:  2001        PMID: 11319876     DOI: 10.1006/jsre.2001.6094

Source DB:  PubMed          Journal:  J Surg Res        ISSN: 0022-4804            Impact factor:   2.192


  3 in total

1.  [Efficacy and safety of various energy sources and application techniques for the surgical treatment of atrial fibrillation].

Authors:  N Doll; H Aupperle; M Borger; M Czesla; F W Mohr
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2007-06

Review 2.  Surgical ablation devices for atrial fibrillation.

Authors:  Shelly C Lall; Ralph J Damiano
Journal:  J Interv Card Electrophysiol       Date:  2007-12       Impact factor: 1.900

Review 3.  Surgical Ablation of Atrial Fibrillation Using Energy Sources.

Authors:  Alexandre Visconti Brick; Domingo Marcolino Braile
Journal:  Braz J Cardiovasc Surg       Date:  2015 Nov-Dec
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.