Literature DB >> 11309003

Computer control versus manual control of systemic hypertension during cardiac surgery.

S A Hoeksel1, J A Blom, J R Jansen, J G Maessen, J J Schreuder.   

Abstract

BACKGROUND: We recently demonstrated the feasibility of computer controlled infusion of vasoactive drugs for the control of systemic hypertension during cardiac surgery. The objective of the current study was to investigate the effects of computer controlled blood pressures on hemodynamic stability when compared to conventional manual control.
METHOD: Systemic artery blood pressures were managed either by computer (80 patients) or by a well-trained anesthesiologist (80 patients). The vasodilator drugs sodium nitroprusside and nitroglycerin were used. Hemodynamic stability was determined from the standard deviation of the mean arterial pressure samples and from the percentages of time that arterial pressure was hypertensive or hypotensive.
RESULTS: The average standard deviation of the mean arterial pressure samples was smaller for the computer controlled than for the manually controlled group: 7.5+/-2.2 (mean+/-SD) versus 8.9+/-2.3 mmHg (P<0.0001). The systemic artery pressure was less hypertensive and less hypotensive in the computer controlled than in the manually controlled group: 9.4+/-5.7 versus 13.1+/-6.0% (P<0.0001) and 8.0+/-5.9 versus 11.8+/-7.4% (P<0.0001), respectively.
CONCLUSION: We conclude that, compared with manual control, computer control of systemic hypertension significantly improved hemodynamic stability during cardiac surgery.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11309003     DOI: 10.1034/j.1399-6576.2001.045005553.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  2 in total

1.  New technology in anaesthesia: friend or foe?

Authors:  R Ross Kennedy
Journal:  J Clin Monit Comput       Date:  2014-04       Impact factor: 2.502

2.  IMPROVED TRANSFORMATION OF MORPHOMETRIC MEASUREMENTS FOR A PRIORI PARAMETER ESTIMATION IN A PHYSIOLOGICALLY-BASED PHARMACOKINETIC MODEL OF ETHANOL.

Authors:  Martin H Plawecki; Ray Decarlo; Vijay A Ramchandani; Sean O'Connor
Journal:  Biomed Signal Process Control       Date:  2007-04       Impact factor: 3.880

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.