Literature DB >> 11308156

Intermediate-term results with 1,019 carbomedics aortic valves.

H H Li1, J Hahn, P Urbanski, M Torka, G L Grunkemeier, R W Hacker.   

Abstract

BACKGROUND: A retrospective study was conducted to evaluate the intermediate-term outcome in patients with the Carbomedics aortic valve prosthesis.
METHODS: The study included 1,019 primary valve replacements between 1989 and 1997. Seventy-two percent of patients were men; mean (standard deviation) age was 61 (10) years. The preoperative New York Heart Association functional class was III or IV in 70% of patients. Follow-up at 9 years was 99.6% complete, comprising 2,730 patient-years (mean, 2.7 years).
RESULTS: Patient survival, including operative deaths, was 80% at 7 years. The linearized death rate was 2.9%/year. Statistically significant risk factors for mortality were diabetes, pure valve insufficiency, advanced age at operation, and advanced preoperative functional class. Linearized rates were thrombosis, 0.1%/year; thromboembolism, 1.0%/year; hemorrhage, 1.7%/year; endocarditis, 0.1%/year; paravalvular leak, 0.1%/year; reoperation, 0.1%/year; and all events, 3.0%/year. The 7-year estimates of freedom from complications were thrombosis, 99%; thromboembolism, 93%; hemorrhage, 89%; endocarditis, 99%; paravalvular leak, 99.7%; reoperation, 99%; and all events, 82%. No structural valve failure was observed.
CONCLUSIONS: The low incidence of valve-related complications favors the continued use of the Carbomedics valve in the aortic position.

Entities:  

Mesh:

Year:  2001        PMID: 11308156     DOI: 10.1016/s0003-4975(00)02652-7

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  4 in total

1.  The CarboMedics prosthetic heart valve: experience with 180 implants.

Authors:  Tsutomu Saito; Yoshio Misawa; Katsuo Fuse; Hiroaki Konishi
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

2.  Complete detachment of an aortic valve prosthesis 10 years after implantation.

Authors:  P Grubwieser; M Pavlic; D Hoefer; W Rabl
Journal:  Int J Legal Med       Date:  2003-10-22       Impact factor: 2.686

3.  When and how does nonstructural mechanical prosthetic heart valve dysfunction occur?

Authors:  Yoshio Misawa; Tsutomu Saito; Hiroaki Konishi; Shin-ichi Oki; Yuichiro Kaminishi; Hideki Takahashi; Kei Aizawa; Osamu Kamisawa; Morito Kato; Katsuo Fuse
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2003-08

4.  Stentless aortic valve replacement in the young patient: long-term results.

Authors:  Torsten Christ; Herko Grubitzsch; Benjamin Claus; Wolfgang Konertz
Journal:  J Cardiothorac Surg       Date:  2013-04-08       Impact factor: 1.637

  4 in total