Literature DB >> 11298712

Electrosurgery, pacemakers and ICDs: a survey of precautions and complications experienced by cutaneous surgeons.

H M El-Gamal1, R G Dufresne, K Saddler.   

Abstract

BACKGROUND: Minimal information is available in the literature regarding the precautions implemented or complications experienced by cutaneous surgeons when electrosurgery is used in patients with pacemakers or implantable cardioverter-defibrillators (ICDs). The literature pertinent to dermatologists is primarily based on experiences of other surgical specialties and a generally recommended thorough perioperative evaluation.
OBJECTIVE: To determine what precautions are currently taken by cutaneous surgeons in patients with pacemakers or ICDs, and what types of complications have occurred due to electrosurgery in a dermatologic setting.
METHODS: In the winter of 2000, a survey was mailed to 419 U.S.-based members of the American College of Mohs Micrographic Surgery and Cutaneous Oncology (ACMMSCO).
RESULTS: A total of 166 (40%) surveys were returned. Routine precautions included utilizing short bursts of less than 5 seconds (71%), use of minimal power (61%), and avoiding use around the pacemaker or ICD (57%). The types of interference reported were skipped beats (eight patients), reprogramming of a pacemaker (six patients), firing of an ICD (four patients), asystole (three patients), bradycardia (two patients), depleted battery life of a pacemaker (one patient), and an unspecified tachyarrhythmia (one patient). Overall there was a low rate of complications (0.8 cases/100 years of surgical practice), with no reported significant morbidity or mortality. Bipolar forceps were utilized by 19% of respondents and were not associated with any incidences of interference.
CONCLUSIONS: Significant interference to pacemakers or ICDs rarely results from office-based electrosurgery. No clear community practice standards regarding precautions was evident from this survey. The use of bipolar forceps or true electrocautery are the better options when electrosurgey is required. These two modalities may necessitate fewer perioperative precautions than generally recommended, without compromising patient safety.

Entities:  

Mesh:

Year:  2001        PMID: 11298712     DOI: 10.1046/j.1524-4725.2001.00287.x

Source DB:  PubMed          Journal:  Dermatol Surg        ISSN: 1076-0512            Impact factor:   3.398


  7 in total

1.  A continuum thermomechanical model of in vivo electrosurgical heating of hydrated soft biological tissues.

Authors:  Wafaa Karaki; Carlos A Lopez; Diana-Andra Borca-Tasciuc; Suvranu De
Journal:  Int J Heat Mass Transf       Date:  2018-07-14       Impact factor: 5.584

Review 2.  Endoscopy and implantable electronic devices.

Authors:  G D Corbett; P C Buttery; P J Pugh; E A B Cameron
Journal:  Frontline Gastroenterol       Date:  2012-02-05

3.  Digital music players cause interference with interrogation telemetry for pacemakers and implantable cardioverter-defibrillators without affecting device function.

Authors:  Gregory Webster; Ligia Jordao; Maria Martuscello; Tarun Mahajan; Mark E Alexander; Frank Cecchin; John K Triedman; Edward P Walsh; Charles I Berul
Journal:  Heart Rhythm       Date:  2008-03-04       Impact factor: 6.343

Review 4.  Preventing and managing complications in dermatologic surgery: Procedural and postsurgical concerns.

Authors:  Allen G Strickler; Payal Shah; Shirin Bajaj; Richard Mizuguchi; Rajiv I Nijhawan; Mercy Odueyungbo; Anthony Rossi; Désirée Ratner
Journal:  J Am Acad Dermatol       Date:  2021-01-23       Impact factor: 15.487

Review 5.  Dermatologic surgery on the chest wall in patients with a cardiac surgery history: a review of material that may be encountered intraoperatively, including potential complications and suggestions for proceeding safely.

Authors:  Elizabeth Foley Bucher; Andrew Kim; Jason Givan; Mary E Maloney
Journal:  Int J Womens Dermatol       Date:  2016-02-28

6.  Management of Facial Telangiectasias with Hand Cautery.

Authors:  Ioannis E Liapakis; Miriam Englander; Roven Sinani; Eleftherios I Paschalis
Journal:  World J Plast Surg       Date:  2015-07

7.  Implantable cardioverter defibrillator during laser transurethral resection of the prostate.

Authors:  A F Deroee; B J Cohen; J F O'Hara
Journal:  Heart Lung Vessel       Date:  2014
  7 in total