A comparison of para-anastomotic compliance profiles after vascular anastomosis: nonpenetrating clips versus standard sutures.

M S Baguneid1, S Goldner, P E Fulford, G Hamilton, M G Walker, A M Seifalian.   

Abstract

PURPOSE: Anastomotic compliance is an important predictive factor for long-term patency of small diameter vascular reconstruction. In this experimental study we compare the compliance of continuous and interrupted sutured vascular anastomoses with those using nonpenetrating clips.
METHODS: Both common carotid arteries in nine goats (average weight, 57 +/- 5.7 kg) were transected, and end-to-end anastomoses were constructed with nonpenetrating clips or polypropylene sutures. The latter were applied with both interrupted and continuous techniques. Intraluminal pressure was measured with a Millar Mikro-tip transducer, and vessel wall motion was determined with duplex ultrasound equipped with an echo-locked wall-tracking system. Diametrical compliance was determined. Environmental scanning electron microscopy was performed on explanted anastomoses.
RESULTS: There was a reduction in anastomotic compliance and associated proximal and distal para-anastomotic hypercompliant zones with the use of all techniques. However, compliance loss was significantly less in those anastomoses with clips and interrupted sutures when compared with continuous suture (P <.001). Furthermore, the total compliance mismatch across anastomoses with continuous sutures was significantly greater than those with clips or interrupted sutures (P <.05). The mean time for constructing clipped anastomoses was 5.7 +/- 1.4 minutes, which was significantly less than either continuous (P <.0001) or interrupted sutures (P <.0001). Furthermore, environmental scanning electron microscopy demonstrated minimal intimal damage with good intimal apposition in the clip group.
CONCLUSION: Anastomoses performed with nonpenetrating clips resulted in improved para-anastomotic compliance profiles and reduced intimal damage when compared with those with polypropylene sutures. These benefits may enhance long-term graft patency by reducing the risk of anastomotic intimal hyperplasia.

Entities:  

Mesh:

Year:  2001        PMID: 11296337     DOI: 10.1067/mva.2001.112806

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  4 in total

Review 1.  Stem cell sources for vascular tissue engineering and regeneration.

Authors:  Vivek K Bajpai; Stelios T Andreadis
Journal:  Tissue Eng Part B Rev       Date:  2012-07-03       Impact factor: 6.389

2.  Biomechanics and biocompatibility of the perfect conduit-can we build one?

Authors:  Michael J Byrom; Martin K C Ng; Paul G Bannon
Journal:  Ann Cardiothorac Surg       Date:  2013-07

3.  Outcomes of Percutaneous Portal Vein Intervention in a Single UK Paediatric Liver Transplantation Programme.

Authors:  Ravi Patel; Jeevan Mahaveer; Nasim Tahir; Sanjay Rajwal; Patricia McClean; Jai V Patel
Journal:  Cardiovasc Intervent Radiol       Date:  2017-09-14       Impact factor: 2.740

4.  Histological Reactions and the In Vivo Patency Rates of Small Silk Vascular Grafts in a Canine Model.

Authors:  Makoto Haga; Satoshi Yamamoto; Hiroyuki Okamoto; Katsuyuki Hoshina; Tetsuro Asakura; Toshiaki Watanabe
Journal:  Ann Vasc Dis       Date:  2017-06-25
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.