Conversion of umbilical arterial Doppler waveforms to cardiac cycle triggering signals: a preparatory study for online motion-gated three-dimensional fetal echocardiography.

J Deng1, A G Birkett, K D Kalache, M A Hanson, D M Peebles, A D Linney, W R Lees, C H Rodeck.   

Abstract

To remove motion artefacts, a device was built to convert "noisy" umbilical arterial Doppler waveforms (UADWs) from an ultrasound (US) system into sharp ECG R-wave-like cardiac cycle triggering signals (CCTSs). These CCTSs were then used to gate a simultaneous (online) 3-D acquisition of sectional fetal echocardiograms from another US system. To test the conversion performance, a study was carried out in sheep fetal twins. Pulmonary arterial flow waveforms (PAFWs) from implanted probes were traced, in the meantime, to determine the reference cardiac cycle. Interference caused by running the two nonsynchronised US systems was controlled to three degrees (not-noticeable, moderate, and severe), together with high (> or = 40 cm/s) and low (< 40) flow velocities on UADWs. The conversion efficiency, assessed by the percentage of UADWs converted into CCTSs, was in the range of 83% to 100% for not-noticeable and moderate interference, and 0% to 71% for severe interference. The triggering accuracy, assessed by [(time lag mean between the onsets of PAFWs and corresponding CCTSs) -- (its 99% confidence level)] / the mean, was 90% to 96% for the not-noticeable interference high- and low-flow groups and for the moderate interference high-flow group; 19% to 93% for the moderate interference low-flow group; and from not obtainable up to 90% for the severe interference groups. The results show that UADWs can be used as a satisfactory online motion-gating source even in the presence of moderate interference. The major problems are from severe interference or moderate interference with low-flow velocity, which can be minimised/eliminated by the integration of the individual systems involved.

Entities:  

Mesh:

Year:  2001        PMID: 11295270     DOI: 10.1016/s0301-5629(00)00315-x

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  6 in total

Review 1.  Three- and 4-dimensional ultrasound in obstetric practice: does it help?

Authors:  Luís F Gonçalves; Wesley Lee; Jimmy Espinoza; Roberto Romero
Journal:  J Ultrasound Med       Date:  2005-12       Impact factor: 2.153

2.  Usefulness of fetal three-dimensional ultrasonography for detecting of congenital heart defects and associated syndromes.

Authors:  Nadia Zabadneh; Claudia Santagati; Elena Reffo; Roberta Biffanti; Alessia Cerutti; Nicola Maschietto; Ornella Milanesi
Journal:  Pediatr Cardiol       Date:  2011-04-09       Impact factor: 1.655

Review 3.  Quantitative in vivo imaging of embryonic development: opportunities and challenges.

Authors:  Chelsea L Gregg; Jonathan T Butcher
Journal:  Differentiation       Date:  2012-06-12       Impact factor: 3.880

4.  Doppler-derived trigger signals for high-frame-rate mouse cardiovascular imaging.

Authors:  Orlando Aristizabal; Jonathan Mamou; Daniel H Turnbull; Jeffrey A Ketterling
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2009

5.  Prospective ECG-gated mouse cardiac imaging with a 34-MHz annular array transducer.

Authors:  Jeffrey A Ketterling; Orlando Aristizábal
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2009-07       Impact factor: 2.725

Review 6.  Translational paradigms in scientific and clinical imaging of cardiac development.

Authors:  Chelsea L Gregg; Jonathan T Butcher
Journal:  Birth Defects Res C Embryo Today       Date:  2013-06
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.