Literature DB >> 11285609

Torn-off balloon tip of Z-5 atrial septostomy catheter.

T Akagi1, Y Tananari, Y V Maeno, W Himeno, J Furui, M Ishii, H Kato.   

Abstract

Creation of atrial communication was performed in a newborn with critical aortic stenosis. After the success of the initial creation, balloon atrial septostomy using Z-5 catheter was performed. When catheter was pulled back, the tip of the balloon was torn off. This experience could be considered as noteworthy when using this catheter in patients with unusually thick atrial septum.

Entities:  

Mesh:

Year:  2001        PMID: 11285609     DOI: 10.1002/ccd.1112

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692