Magnet design with high B(0) homogeneity for fast-field-cycling NMR applications.

O Lips1, A F Privalov, S V Dvinskikh, F Fujara.   

Abstract

The design, construction, and performance of a low-inductance solenoidal coil with high B(0) homogeneity for fast-field-cycling NMR is presented. It consists of six concentric layers. The conductor width is varied to minimize the B(0) inhomogeneity in the volume of the sample. This is done using an algorithm which takes the real shape of the conductor directly into account. The calculated coil geometry can be manufactured easily using standard computerized numeric control equipment, which keeps the costs low. The coil is liquid cooled and produces a B(0) field of 0.95 T at 800 A. The field inhomogeneity in a cylindrical volume (diameter 5 mm, length 10 mm) is about 10 ppm, and the inductance is 190 microH. Switching times below 200 micros can be achieved. During 6 months of operation the coil has shown good stability and reliability. Copyright 2001 Academic Press.

Mesh:

Year:  2001        PMID: 11273747     DOI: 10.1006/jmre.2000.2279

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  1 in total

Review 1.  Approaches in cooling of resistive coil-based low-field Magnetic Resonance Imaging (MRI) systems for application in low resource settings.

Authors:  Faith Natukunda; Theodora M Twongyirwe; Steven J Schiff; Johnes Obungoloch
Journal:  BMC Biomed Eng       Date:  2021-02-12
  1 in total