Clinical use of WHO haemoglobin colour scale: validation and critique.

C F Ingram1, S M Lewis.   

Abstract

AIM: The World Health Organisation (WHO) haemoglobin colour scale has been developed as a simple, inexpensive clinical device for diagnosing anaemia when laboratory based haemoglobinometry is not available. In an initial validation study at several health centres, scale readings were compared with measurements of haemoglobin by the laboratory. This showed the scale to have 90% sensitivity and 70% specificity in identifying whether anaemia was present or not. In addition, when present, the degree of anaemia was correctly classified in clinical terms as moderate, pronounced, or severe, with an overall sensitivity of 60% and specificity of 88%. Errors were mainly marginal--that is, between two adjacent categories--but there were also some major discrepancies, such as a blood with a haemoglobin of 6-7 g/dl being read as normal or vice versa. Because this would compromise the scale's reliability in practice, this study was undertaken to identify the cause of the discrepancies and to reassess the performance of the scale.
METHODS: Venous blood samples were collected into potassium EDTA from patients attending selected clinics at three South African hospitals with good laboratory facilities. A prototype of the device was used unsupervised by nursing staff, doctors, and phlebotomists, who were told to follow the printed instructions. The blood specimens were then immediately sent to the laboratory where haemoglobin was measured by standardised automated blood cell counters. Any discrepancies > 1 g/dl were recorded and the tests were repeated by the same operators under supervision of the investigators.
RESULTS: Almost all the errors that occurred resulted from the incorrect use of the device, namely: inadequate or excessive blood, reading the results too soon or too late (beyond the limit of two minutes), poor lighting, or holding the scale at the wrong angle. The accuracy improved dramatically when the tests were repeated under supervision and these faults were avoided: 95% of readings were within 1 g/dl of the reference measurements, and 97% within 1.5 g/dl. Anaemia screening showed 96% sensitivity and 86% specificity. Clinical judgement of pallor was frequently wrong, whereas the scale gave the correct diagnosis in more than 97% of cases.
CONCLUSION: The study confirmed the usefulness and reliability of the scale and its advantage over clinical signs for the diagnosis of anaemia, thus providing a clinically reliable near patient method in the absence of a laboratory. The instructions are easy to follow but must be strictly adhered to.

Entities:  

Mesh:

Year:  2000        PMID: 11265179      PMCID: PMC1731129          DOI: 10.1136/jcp.53.12.933

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  15 in total

1.  Validation of the haemoglobin colour scale for screening blood donors in Malawi.

Authors:  B M'baya; I Mbingwani; L Mgawi; V Mkochi; I Bates; S A White; T J Allain
Journal:  Malawi Med J       Date:  2014-06       Impact factor: 0.875

2.  Disposable platform provides visual and color-based point-of-care anemia self-testing.

Authors:  Erika A Tyburski; Scott E Gillespie; William A Stoy; Robert G Mannino; Alexander J Weiss; Alexa F Siu; Rayford H Bulloch; Karthik Thota; Anyela Cardenas; Wilena Session; Hanna J Khoury; Siobhán O'Connor; Silvia T Bunting; Jeanne Boudreaux; Craig R Forest; Manila Gaddh; Traci Leong; L Andrew Lyon; Wilbur A Lam
Journal:  J Clin Invest       Date:  2014-08-26       Impact factor: 14.808

3.  3D printed auto-mixing chip enables rapid smartphone diagnosis of anemia.

Authors:  Kimberly Plevniak; Matthew Campbell; Timothy Myers; Abby Hodges; Mei He
Journal:  Biomicrofluidics       Date:  2016-10-05       Impact factor: 2.800

4.  Laboratory practice at the periphery in developing countries.

Authors:  S M Lewis
Journal:  Int J Hematol       Date:  2002-08       Impact factor: 2.490

Review 5.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

Review 6.  Iron deficiency anemia: a common and curable disease.

Authors:  Jeffery L Miller
Journal:  Cold Spring Harb Perspect Med       Date:  2013-07-01       Impact factor: 6.915

7.  Quality evaluation of four hemoglobin screening methods in a blood donor setting along with their comparative cost analysis in an Indian scenario.

Authors:  Rashmi Tondon; Anupam Verma; Prashant Pandey; Rajendra Chaudhary
Journal:  Asian J Transfus Sci       Date:  2009-07

8.  Accuracy of clinical pallor in the diagnosis of anaemia in children: a meta-analysis.

Authors:  Juan P Chalco; Luis Huicho; Carlos Alamo; Nilton Y Carreazo; Carlos A Bada
Journal:  BMC Pediatr       Date:  2005-12-08       Impact factor: 2.125

9.  Assessing diagnostic accuracy of Haemoglobin Colour Scale in real-life setting.

Authors:  Pankaj P Shah; Shrey A Desai; Dhiren K Modi; Shobha P Shah
Journal:  J Health Popul Nutr       Date:  2014-03       Impact factor: 2.000

10.  Validity of Palmar Pallor for Diagnosis of Anemia among Children Aged 6-59 Months in North India.

Authors:  Arun Kumar Aggarwal; Jaya Prasad Tripathy; Deepak Sharma; Ajith Prabhu
Journal:  Anemia       Date:  2014-11-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.