Comparison of nebulized particle size distribution with Malvern laser diffraction analyzer versus Andersen cascade impactor and low-flow Marple personal cascade impactor.

W T Kwong1, S L Ho, A L Coates.   

Abstract

Particle size of nebulized aerosols can be measured directly using laser diffraction or by evaluating aerodynamic properties by cascade impaction. As of today, there are no generally accepted standards for measuring particle size distribution from nebulizers. Laser diffraction has been questioned because of potential evaporative losses of the small particles at the edge of the plume, causing an apparent shift in the particle size distribution and thus a larger mass median diameter (MMD). When particle-sizing wet aerosols, cascade impaction may give rise to an apparent shift in the distribution, resulting in a smaller mass median aerodynamic diameter (MMAD) due to evaporative losses of aerosol droplets as they enter the impactor at ambient temperature. The modified low-flow Marple 296 Personal Cascade Impactor (MPCI) is currently being proposed as the European standard for wet aerosol analysis to minimize evaporative losses during sampling. The present study compared the particle size distribution of salbutamol and sodium cromoglycate aerosols nebulized by the Pari LC Star, using laser diffraction (Malvern Mastersizer X; MMX) and cascade impaction (Andersen Cascade Impactor [ACI] and the commercially available MPCI), which was either at ambient temperature or cooled to the nebulized aerosol temperature (10 degrees C). MMDs obtained with the MMX were virtually identical to the MMADs measured with both impactors when cooled with no significant differences in geometric standard deviation (sigma(g)). When the impactors were operated at ambient temperature, MMADs were smaller (18 to 30%) with a significantly larger sigma(g) (p < 0.05) compared to the MMX. These findings suggest that droplet distribution data for wet aerosol where evaporation process has not been minimized must be viewed with caution. There was no evidence suggesting a significant evaporative loss of small droplets from the edge of the plume during laser particle sizing. The MPCI does not minimize evaporative losses of aerosol particles during sampling.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11262437     DOI: 10.1089/jam.2000.13.303

Source DB:  PubMed          Journal:  J Aerosol Med        ISSN: 0894-2684


  6 in total

Review 1.  Aerosol delivery to ventilated newborn infants: historical challenges and new directions.

Authors:  Jan Mazela; Richard A Polin
Journal:  Eur J Pediatr       Date:  2010-09-28       Impact factor: 3.183

Review 2.  Non-impactor-based methods for sizing of aerosols emitted from orally inhaled and nasal drug products (OINDPs).

Authors:  Jolyon Mitchell; Richard Bauer; Svetlana Lyapustina; Terrence Tougas; Volker Glaab
Journal:  AAPS PharmSciTech       Date:  2011-07-22       Impact factor: 3.246

3.  Effects of Temperature and Humidity on Laser Diffraction Measurements to Jet Nebulizer and Comparison with NGI.

Authors:  Xinghan Song; Junhua Hu; Shuyao Zhan; Rui Zhang; Wen Tan
Journal:  AAPS PharmSciTech       Date:  2015-07-14       Impact factor: 3.246

4.  Nebulization of Single-Chain Tissue-Type and Single-Chain Urokinase Plasminogen Activator for Treatment of Inhalational Smoke-Induced Acute Lung Injury.

Authors:  Soraya Hengsawas Surasarang; Sawittree Sahakijpijarn; Galina Florova; Andrey A Komissarov; Christina L Nelson; Enkhbaatar Perenlei; Satoshi Fukuda; Marla R Wolfson; Thomas H Shaffer; Steven Idell; Robert O Williams
Journal:  J Drug Deliv Sci Technol       Date:  2018-04-30       Impact factor: 3.981

5.  Determination of reference ultrasound parameters for model and hydrofluoroalkane propellants using high-resolution ultrasonic spectroscopy.

Authors:  Susan Hoe; Paul M Young; Philippe Rogueda; Daniela Traini
Journal:  AAPS PharmSciTech       Date:  2008-05-06       Impact factor: 3.246

6.  Humidified and Heated Cascade Impactor for Aerosol Sizing.

Authors:  Caroline Majoral; Allan L Coates; Alain Le Pape; Laurent Vecellio
Journal:  Front Bioeng Biotechnol       Date:  2020-11-13
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.