MRI monitoring of the thermal ablation of tissue: effects of long exposure times.

N McDannold1, K Hynynen, F Jolesz.   

Abstract

MRI-derived thermometry based on the temperature-dependence of the proton resonant frequency (PRF) is extremely sensitive to changes in tissue unrelated to temperature changes, including tissue swelling. This study investigated the maximum amount of time that this phase-subtraction-based method can be used to accurately monitor temperature changes in vivo. Long-duration focused ultrasound sonications were delivered in rabbit thigh muscle with a phased-array transducer, and the time that tissue swelling began was monitored. Tissue swelling began to occur at about one minute. The temperature correlated well with an implanted thermocouple up to this time. After this time, severe artifacts in the phase-difference maps were observed. The thermal dose model predicted the extent of tissue damage well for subsequent one minute sonications. These results will have implications for MRI guidance of thermal therapies with long exposure times.

Entities:  

Mesh:

Year:  2001        PMID: 11241817     DOI: 10.1002/jmri.1061

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  14 in total

Review 1.  [Magnetic resonance tomography guided focussed ultrasound surgery (MRgFUS) in tumor therapy--a new noninvasive therapy option].

Authors:  S A Hengst; T Ehrenstein; H Herzog; A Beck; I Utz-Billing; M David; R Felix; J Ricke
Journal:  Radiologe       Date:  2004-04       Impact factor: 0.635

2.  Uterine leiomyomas: MR imaging-based thermometry and thermal dosimetry during focused ultrasound thermal ablation.

Authors:  Nathan McDannold; Clare M Tempany; Fiona M Fennessy; Minna J So; Frank J Rybicki; Elizabeth A Stewart; Ferenc A Jolesz; Kullervo Hynynen
Journal:  Radiology       Date:  2006-07       Impact factor: 11.105

Review 3.  MR thermometry.

Authors:  Viola Rieke; Kim Butts Pauly
Journal:  J Magn Reson Imaging       Date:  2008-02       Impact factor: 4.813

4.  Prevention of post-focal thermal damage by formation of bubbles at the focus during high intensity focused ultrasound therapy.

Authors:  Vesna Zderic; Jessica Foley; Wenbo Luo; Shahram Vaezy
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

Review 5.  MRI-guided focused ultrasound surgery.

Authors:  Ferenc A Jolesz
Journal:  Annu Rev Med       Date:  2009       Impact factor: 13.739

6.  Multiple applicator hepatic ablation with interstitial ultrasound devices: theoretical and experimental investigation.

Authors:  Punit Prakash; Vasant A Salgaonkar; E Clif Burdette; Chris J Diederich
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

7.  Noninvasive assessment of tissue heating during cardiac radiofrequency ablation using MRI thermography.

Authors:  Aravindan Kolandaivelu; Menekhem M Zviman; Valeria Castro; Albert C Lardo; Ronald D Berger; Henry R Halperin
Journal:  Circ Arrhythm Electrophysiol       Date:  2010-07-24

8.  Iterative Curve Fitting of the Bioheat Transfer Equation for Thermocouple-Based Temperature Estimation In Vitro and In Vivo.

Authors:  Hermes A S Kamimura; Christian Aurup; Ethan V Bendau; Niloufar Saharkhiz; Min Gon Kim; Elisa E Konofagou
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2019-09-11       Impact factor: 2.725

9.  MRI-based thermal dosimetry and diffusion-weighted imaging of MRI-guided focused ultrasound thermal ablation of uterine fibroids.

Authors:  Magdalini C Pilatou; Elizabeth A Stewart; Stephan E Maier; Fiona M Fennessy; Kullervo Hynynen; Clare M C Tempany; Nathan McDannold
Journal:  J Magn Reson Imaging       Date:  2009-02       Impact factor: 4.813

10.  Drift correction for accurate PRF-shift MR thermometry during mild hyperthermia treatments with MR-HIFU.

Authors:  Chenchen Bing; Robert M Staruch; Matti Tillander; Max O Köhler; Charles Mougenot; Mika Ylihautala; Theodore W Laetsch; Rajiv Chopra
Journal:  Int J Hyperthermia       Date:  2016-05-22       Impact factor: 3.914

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.