Literature DB >> 11241515

Cryogen spray cooling: Effects of droplet size and spray density on heat removal.

B M Pikkula1, J H Torres, J W Tunnell, B Anvari.   

Abstract

BACKGROUND AND
OBJECTIVE: Cryogen spray cooling (CSC) is an effective method to reduce or eliminate non-specific injury to the epidermis during laser treatment of various dermatological disorders. In previous CSC investigations, fuel injectors have been used to deliver the cryogen onto the skin surface. The objective of this study was to examine cryogen atomization and heat removal characteristics of various cryogen delivery devices. STUDY DESIGN/
MATERIALS AND METHODS: Various cryogen delivery device types including fuel injectors, atomizers, and a device currently used in clinical settings were investigated. Cryogen mass was measured at the delivery device output orifice. Cryogen droplet size profiling for various cryogen delivery devices was estimated by optically imaging the droplets in flight. Heat removal for various cryogen delivery devices was estimated over a range of spraying distances by temperature measurements in an skin phantom used in conjunction with an inverse heat conduction model.
RESULTS: A substantial range of mass outputs were measured for the cryogen delivery devices while heat removal varied by less than a factor of two. Droplet profiling demonstrated differences in droplet size and spray density.
CONCLUSIONS: Results of this study show that variation in heat removal by different cryogen delivery devices is modest despite the relatively large difference in cryogen mass output and droplet size. A non-linear relationship between heat removal by various devices and droplet size and spray density was observed. Copyright 2001 Wiley-Liss, Inc.

Mesh:

Substances:

Year:  2001        PMID: 11241515     DOI: 10.1002/lsm.1024

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  2 in total

1.  [Local anesthetic procedures in dermatology. Part 2: Practical aspects].

Authors:  D Dill-Müller
Journal:  Hautarzt       Date:  2012-02       Impact factor: 0.751

2.  Heat exchange between a bouncing drop and a superhydrophobic substrate.

Authors:  Samira Shiri; James C Bird
Journal:  Proc Natl Acad Sci U S A       Date:  2017-06-19       Impact factor: 11.205

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.