Literature DB >> 11236873

Sensitivity maps and system requirements for magnetic induction tomography using a planar gradiometer.

J Rosell1, R Casañas, H Scharfetter.   

Abstract

We evaluated analytically and experimentally the performance of a planar gradiometer as a sensing element in a system for magnetic induction tomography. A system using an excitation coil and a planar gradiometer was compared against a system with two coils. We constructed one excitation coil, two different sensing elements and a high-resolution phase detector. The first sensor was a PCB square spiral coil with seven turns. The second sensor was a PCB planar gradiometer with two opposite square spirals of seven turns, with a distance between centres of 8 cm. Theoretical sensitivity maps were derived from basic equations and compared with experimental data obtained at 150 kHz. The experimental sensitivity maps were obtained measuring the perturbation produced by a brass sphere of 12 mm in empty space. The advantage of using a gradiometer is that it can be adjusted to give a minimum signal for homogeneous objects, while increasing the sensitivity to local perturbations of the conductivity. Results show that a system using a planar gradiometer as detector has less demanding requirements for the electronic system than a system using simple coils.

Entities:  

Mesh:

Year:  2001        PMID: 11236873     DOI: 10.1088/0967-3334/22/1/316

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  5 in total

1.  Vector projection of biomagnetic fields.

Authors:  L A Bradshaw; A Myers; W O Richards; W Drake; J P Wikswo
Journal:  Med Biol Eng Comput       Date:  2005-01       Impact factor: 2.602

2.  Single-step 3-d image reconstruction in magnetic induction tomography: theoretical limits of spatial resolution and contrast to noise ratio.

Authors:  Hermann Scharfetter; Karl Hollaus; Javier Rosell-Ferrer; Robert Merwa
Journal:  Ann Biomed Eng       Date:  2006-10-10       Impact factor: 3.934

3.  Extraction of single-trial cortical beta oscillatory activities in EEG signals using empirical mode decomposition.

Authors:  Chia-Lung Yeh; Hsiang-Chih Chang; Chi-Hsun Wu; Po-Lei Lee
Journal:  Biomed Eng Online       Date:  2010-06-17       Impact factor: 2.819

Review 4.  Advancements in transmitters and sensors for biological tissue imaging in magnetic induction tomography.

Authors:  Zulkarnay Zakaria; Ruzairi Abdul Rahim; Muhammad Saiful Badri Mansor; Sazali Yaacob; Nor Muzakkir Nor Ayub; Siti Zarina Mohd Muji; Mohd Hafiz Fazalul Rahiman; Syed Mustafa Kamal Syed Aman
Journal:  Sensors (Basel)       Date:  2012-05-29       Impact factor: 3.576

Review 5.  Screening and Biosensor-Based Approaches for Lung Cancer Detection.

Authors:  Lulu Wang
Journal:  Sensors (Basel)       Date:  2017-10-23       Impact factor: 3.576

  5 in total