Hemodynamic evaluation of 19-mm Carpentier-Edwards pericardial bioprosthesis in aortic position.

H Takakura1, T Sasaki, K Hashimoto, T Hachiya, K Onoguchi, M Oshiumi, S Takeuchi.   

Abstract

BACKGROUND: The aortic Carpentier-Edwards pericardial bioprosthesis offers good long-term clinical outcomes with a low rate of structural deterioration. However, little in vivo hemodynamic data is available for this bioprosthesis.
METHODS: To determine the hemodynamic performance of the 19-mm Carpentier-Edwards pericardial valve, both cardiac catheterization and dobutamine stress echocardiography were electively performed in 10 patients. The mean age at the study was 71.6 +/- 4.4 years and the mean body surface area was 1.39 +/- 0.11 m2. The peak-to-peak gradient, instantaneous peak gradient, mean gradient, and valve orifice area were measured by standard cardiac catheterization. The Doppler-derived gradients and valve orifice area were also measured both at rest and during dobutamine infusion.
RESULTS: The average peak-to-peak gradient, instantaneous peak gradient, mean gradient, and valve orifice area measured by catheterization were 13.0 +/- 5.4 mmHg, 28.5 +/- 7.7 mmHg, 12.0 +/- 4.9 mmHg, and 1.55 +/- 0.45 cm2, respectively. The peak and mean Doppler gradients, and valve orifice area by resting echocardiography were 27.7 +/- 9.5 mmHg, 12.3 +/- 4.8 mmHg, and 1.39 +/- 0.26 cm2, respectively. At a dosage of 10 microg/kg/min of dobutamine, the mean Doppler gradient rose mildly to 22.2 +/- 4.8 mmHg, while the cardiac output increased from 4.49 +/- 0.44 to 6.64 +/- 0.87 L/min. The valve orifice area during the 10 microg/kg/min dobutamine infusion (1.55 +/- 0.25 cm2) was significantly larger than its value at rest (p < 0.05).
CONCLUSIONS: With acceptable hemodynamic performance, use of the aortic 19-mm Carpentier-Edwards pericardial valve is a reliable option for elderly patients with a small annulus.

Entities:  

Mesh:

Year:  2001        PMID: 11235715     DOI: 10.1016/s0003-4975(00)02210-4

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

Review 1.  Update on aortic valve prosthesis-patient mismatch in Japan.

Authors:  Yoshimasa Sakamoto; Kazuhiro Hashimoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-04-13

2.  Patient-prosthesis mismatch may be irrelevant after aortic valve replacement with the 19-mm Perimount pericardial bioprosthesis in patients aged 65 years or older.

Authors:  Yoshimasa Sakamoto; Kazuhiro Hashimoto; Hiroshi Okuyama; Tatsuumi Sasaki; Hiromitsu Takakura; Katsuhisa Onoguchi
Journal:  J Artif Organs       Date:  2007-12-20       Impact factor: 1.731

3.  Early and mid-term outcome in terms of functional and hemodynamic performance of the st. Jude regent 19-mm aortic mechanical prosthesis versus 19-mm carpentier edwards aortic biological prosthesis.

Authors:  Edvin Prifti; Massimo Bonacchi; Fadil Ademaj; Gabriele Giunti; Giampiero Esposito; Arben Baboci; Gani Bajraktari; Altin Veshti; Aurel Demiraj; Vittorio Vanini
Journal:  J Cardiothorac Surg       Date:  2015-11-06       Impact factor: 1.637

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.