First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator. Advantages and complications. The Ventak AV II DR investigators.

C Sticherling1, A Schaumann, T Klingenheben, S H Hohnloser.   

Abstract

AIMS: The need for physiological pacing and for improving the ability to discriminate atrial from ventricular tachyarrhythmias has prompted the development of dual chamber implantable cardioverter/defibrillators (ICDs).
METHODS: Fifty-two patients were implanted with a newly developed dual-chamber ICD providing rate-responsive physiological pacing (Ventak AV II DR). The device possesses two new arrhythmia detection algorithms ('atrial fibrillation rate threshold' and 'ventricular to atrial rate relationship') in addition to commonly used features such as 'onset' and 'stability'. During implantation, the atrial and ventricular lead impedances and pacing thresholds were determined together with the defibrillation threshold. Prior to discharge, attempts were made to induce both atrial and ventricular tachyarrhythmias in order to test those new detection criteria. All patients were followed for at least 3 months.
RESULTS: The device was successfully implanted in all 52 patients. Placement of the atrial lead was successful in 50/52 patients (96%; P-wave 3.2 +/- 1.4 mV; impedance 576 +/- 123 omega; atrial pacing threshold 1.2 +/- 0.9 V). Prior to discharge, 32 episodes of atrial fibrillation (AF) alone, 38 episodes of AF with ventricular fibrillation and 10 episodes of AF with monomorphic ventricular tachycardia were induced in 33/50 patients (66%) and all were appropriately classified by the detection algorithm. During the 3 months follow-up, 12 patients (23%) had appropriate and successful therapies for ventricular arrhythmias, while four patients (8%) experienced inappropriate ICD therapies. Although all these episodes were detected correctly as supraventricular arrhythmias by the device, therapy was delivered because of incorrect or incomplete programming. In all cases reprogramming of the device resolved the problem.
CONCLUSION: Implantation of dual chamber ICDs is feasible and appears to improve discrimination of supraventricular from ventricular tachyarrhythmias. In addition, patients with tachyarrhythmias and concomitant bradyarrhythmias may benefit from simultaneous physiological pacing. However, implantation and follow-up of such patients should be performed at experienced centres since both surgical handling and programming of these devices is more difficult and complex than conventional ICDs.

Entities:  

Mesh:

Year:  1999        PMID: 11233190     DOI: 10.1053/eupc.1998.0023

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

Review 1.  Algorithms for better arrhythmia discrimination in implantable cardioverter defibrillators.

Authors:  R Schimpf; C Wolpert; B Lüderitz
Journal:  Curr Cardiol Rep       Date:  2001-11       Impact factor: 2.931

Review 2.  Prevalence and incidence of arrhythmias and sudden death in heart failure.

Authors:  John G F Cleland; Sudipta Chattopadhyay; Aleem Khand; Timothy Houghton; Gerald C Kaye
Journal:  Heart Fail Rev       Date:  2002-07       Impact factor: 4.214

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.