The new hematology analyzer Sysmex XE-2100: performance evaluation of a novel white blood cell differential technology.

K Ruzicka1, M Veitl, R Thalhammer-Scherrer, I Schwarzinger.   

Abstract

CONTEXT: The new hematology analyzer Sysmex XE-2100 (TOA Medical Electronics, Kobe, Japan) has a novel, combined, white blood cell differential technology and a special reagent system to enumerate nucleated red blood cells.
DESIGN: Performance evaluation of both technologies of the Sysmex XE-2100 according to the H20-A protocol of the National Committee for Clinical and Laboratory Standards and comparison of the results with those for the hematology analyzer Sysmex NE-8000 (TOA Medical Electronics). SPECIMENS: Five hundred forty-four blood samples randomly chosen from various inpatient and outpatient departments of the Vienna University hospital.
RESULTS: Five-part white blood cell differential counts on the XE-2100 revealed excellent correlation with the manual reference method for neutrophils, lymphocytes, and eosinophils (r =.925,.922, and.877, respectively) and good correlation for monocytes and basophils (r =.756 and.763, respectively). The efficiency rates of flagging for the presence of >/=1% abnormal white blood cells were 83% (XE-2100) and 66% (NE-8000). The correlation of automated and microscopic nucleated red blood cell counts was excellent (r =.97).
CONCLUSIONS: From the present evaluation and our former experience with other types of Sysmex analyzers, we conclude that the new white blood cell differential technology of the XE-2100 represents a further development toward more efficient flagging of abnormal white blood cells.

Entities:  

Mesh:

Year:  2001        PMID: 11231489     DOI: 10.5858/2001-125-0391-TNHASX

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  27 in total

1.  Evaluation of the Sysmex Xe-2100 hematology analyzer in hospital use.

Authors:  Danièle Nakul-Aquaronne; Isabelle Sudaka-Sammarcelli; Corinne Ferrero-Vacher; Benoit Starck; Jacques Bayle
Journal:  J Clin Lab Anal       Date:  2003       Impact factor: 2.352

2.  Automatic analysis of normal bone marrow blood cells using the XE-2100 automated hematology analyzer.

Authors:  Hisako Shibata; Takahisa Yamane; Ryousuke Yamamura; Kensuke Ohta; Takayuki Takubo; Tomio Kamitani; Masayuki Hino
Journal:  J Clin Lab Anal       Date:  2003       Impact factor: 2.352

3.  Useful information provided by graphic displays of automated cell counter in hematological malignancies.

Authors:  Monica Gupta; Kriti Chauhan; Tanvi Singhvi; Manisha Kumari; Rajesh Kumar Grover
Journal:  J Clin Lab Anal       Date:  2018-01-21       Impact factor: 2.352

4.  Design of malaria diagnostic criteria for the Sysmex XE-2100 hematology analyzer.

Authors:  Germán Campuzano-Zuluaga; Gonzalo Alvarez-Sánchez; Gloria Elcy Escobar-Gallo; Luz Marina Valencia-Zuluaga; Alexandra Marcela Ríos-Orrego; Adriana Pabón-Vidal; Andrés Felipe Miranda-Arboleda; Silvia Blair-Trujillo; Germán Campuzano-Maya
Journal:  Am J Trop Med Hyg       Date:  2010-03       Impact factor: 2.345

5.  Utility of immature granulocyte percentage in pediatric appendicitis.

Authors:  Eleanor K Mathews; Russell L Griffin; Vincent Mortellaro; Elizabeth A Beierle; Carroll M Harmon; Mike K Chen; Robert T Russell
Journal:  J Surg Res       Date:  2014-04-12       Impact factor: 2.192

6.  Impact of spinach consumption on DNA stability in peripheral lymphocytes and on biochemical blood parameters: results of a human intervention trial.

Authors:  Beate Moser; Thomas Szekeres; Christian Bieglmayer; Karl-Heinz Wagner; Miroslav Mišík; Michael Kundi; Oliwia Zakerska; Armen Nersesyan; Nina Kager; Johann Zahrl; Christine Hoelzl; Veronika Ehrlich; Siegfried Knasmueller
Journal:  Eur J Nutr       Date:  2011-03-09       Impact factor: 5.614

7.  A flow cytometric method for determination of the blood neutrophil fraction in rats.

Authors:  Spela Skrajnar; Marija Anzur Lasnik; Apolonija Bedina Zavec
Journal:  J Am Assoc Lab Anim Sci       Date:  2009-03       Impact factor: 1.232

8.  5'-AMP impacts lymphocyte recirculation through activation of A2B receptors.

Authors:  Hjalmar R Bouma; Judith N Mandl; Arjen M Strijkstra; Ate S Boerema; Jan-Willem Kok; Annie van Dam; Ad Ijzerman; Frans G M Kroese; Robert H Henning
Journal:  J Leukoc Biol       Date:  2013-05-16       Impact factor: 4.962

9.  Choosing wisely during the COVID-19 pandemic: optimising outpatient cancer care while conserving resources with a new algorithm to report automated ANC results.

Authors:  Maly Fenelus; Tamiqua Graham; Ryan Golden; Jessica L Bautista; Rachel J So; Nora Plante; Ellinor I B Peerschke
Journal:  J Clin Pathol       Date:  2020-11-16       Impact factor: 3.411

Review 10.  Automated haematology analysis to diagnose malaria.

Authors:  Germán Campuzano-Zuluaga; Thomas Hänscheid; Martin P Grobusch
Journal:  Malar J       Date:  2010-11-30       Impact factor: 2.979

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.