An electronic device for needle placement during sonographically guided percutaneous intervention.

M H Howard1, R C Nelson, E K Paulson, M A Kliewer, D H Sheafor.   

Abstract

An electronic device for guiding needle placement during sonographically directed percutaneous intervention was tested in a phantom and then in patients. In the phantom, targeting accuracy was similar for use of the needle guide alone, the needle guide with the device, and freehand techniques with the device, but all were superior to the freehand technique alone (P <.001). In 34 (79%) of 43 patients, the device worked well.

Entities:  

Mesh:

Year:  2001        PMID: 11230674     DOI: 10.1148/radiology.218.3.r01mr10905

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105