Literature DB >> 11225875

Determination of uranium in tap water by ICP-MS.

M El Himri1, A Pastor, M de la Guardia.   

Abstract

A fast and accurate procedure has been developed for the determination of uranium at microg L(-1) level in tap and mineral water. The method is based on the direct introduction of samples, without any chemical pre-treatment, into an inductively coupled plasma mass spectrometer (ICP-MS). Uranium was determined at the mass number 238 using Rh as internal standard. The method provides a limit of detection of 2 ng L(-1) and a good repeatability with relative standard deviation values (RSD) about 3% for five independent analyses of samples containing 73 microg L(-1) of uranium. Recovery percentage values found for the determination of uranium in spiked natural samples varied between 91% and 106%. Results obtained are comparable with those found by radiochemical methods for natural samples and of the same order for the certified content of a reference material, thus indicating the accuracy of the ICP-MS procedure without the need of using isotope dilution. A series of mineral and tap waters from different parts of Spain and Morocco were analysed.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11225875     DOI: 10.1007/s002160051616

Source DB:  PubMed          Journal:  Fresenius J Anal Chem        ISSN: 0937-0633